使用非接触式光学生物测量仪LenStar LS 900和IOL Master V.5获得的生物测量数据的比较。

Comparison of the biometric measurements obtained using noncontact optical biometers LenStar LS 900 and IOL Master V.5.

作者信息

Kołodziejczyk Wojciech, Gałecki Tomasz, Łazicka-Gałecka Monika, Szaflik Jerzy

机构信息

Department of Ophthalmology, Medical University of Warsaw, Poland.

出版信息

Klin Oczna. 2011;113(1-3):47-51.

Abstract

PURPOSE

Evaluation of the measurement accuracy using a new optical biometer LenStar LS 900 (Haag Strait) and comparison according to the IOL Master V.5 (Zeiss) device.

METHODS

In a prospective clinical study biometric measurements along with artificial lens power calculation using the LenStar LS 900 device were performed. A total number of 106 patients qualified for a cataract extraction procedure were included in the study. Measurements along with lens power calculation were repeated using the IOL Master V.5. device. Results were elaborated using Pearson's correlation and Bland-Altman plot.

RESULTS

Measurements were performed in 204 eyes of 106 patients. Mean values and ranges of biometry results were as follows: 23.46 +/- 2.81 mm [20.79-29.80] for IOL Master, and 23.47 +/- 2.83 mm [20.79-29.86] for LenStar LS 900. In keratometry for flat meridian (K1) the following data was obtained: mean 43.24 +/- 3.22 D [range 38.27-47.94] for IOL Master and 42.44 +/- 3.15 D [38.27-47.94] for LenStar LS 900. For steep meridian (K2) the data obtained were 44.14 +/- 3.40 D; [39.29-49.13] and 43.27 +/- 3.34 D [38.61-48.4] respectively. The obtained calculation results for each eye were as follows: mean 21.23 +/- 8.07, [range 3.25-28.99] for Hagis formula, 21.14 +/- 6.90; [4.83-27.6] for SRK II, 21.04 +/- 7.78; [3.05-28.05] for SRK/T, 21.09 +/- 8.13; [2.43-28.61] for Holladay using IOL Master and 21.41 +/- 8.23; [2.99-29.15] for Haigis; 21.24 +/- 7.00; [4.6-27.71] for SRK II, 21.13 +/- 7.90; [2.76-28.18] for SRK/T; 21.09 +/- 8.13; [2.16-28.76] for Holladay using LenStar LS 900.

CONCLUSIONS

The LenStar LS 900 device enables to perform accurate and repetitive biometric measurements and implant power calculations. Implant calculation results obtained using the LenStar LS 900 device are comparable to those achieved using the IOL Master V.5 device, which has been commonly accepted as standard for over a decade. The use of both devices is limited by significant lens opacification and posterior capsule calcification. In such cases, additional ultrasound biometry should be performed. Keratometry results obtained using both devices should not be used alternatively because of the different measurement methods and different refraction indexes. The LenStar LS 900 device comparing to the IOL Master additionally enables pachymetry, macular retinal thickness, lens thickens and pupil diameter measurement. Accuracy of those measurements should be studied in the future.

摘要

目的

使用新型光学生物测量仪LenStar LS 900( Haag Strait)评估测量准确性,并与IOL Master V.5(蔡司)设备进行比较。

方法

在一项前瞻性临床研究中,使用LenStar LS 900设备进行生物测量以及人工晶状体度数计算。共有106例符合白内障摘除手术条件的患者纳入研究。使用IOL Master V.5设备重复进行测量以及晶状体度数计算。使用Pearson相关性分析和Bland-Altman图对结果进行阐述。

结果

对106例患者的204只眼进行了测量。生物测量结果的平均值和范围如下:IOL Master测量的眼轴长度为23.46±2.81mm[20.79 - 29.80],LenStar LS 900测量的为23.47±2.83mm[20.79 - 29.86]。在角膜曲率计测量的平轴子午线(K1)中,获得的数据如下:IOL Master测量的平均值为43.24±3.22D[范围38.27 - 47.94],LenStar LS 900测量的为42.44±3.15D[38.27 - 47.94]。对于陡轴子午线(K2),获得的数据分别为44.14±3.40D;[39.29 - 49.13]和43.27±3.34D[38.61 - 48.4]。每只眼获得的计算结果如下:使用IOL Master通过Hagis公式计算的平均值为21.23±8.07,[范围3.25 - 28.99],SRK II公式计算的为21.14±6.90;[4.83 - 27.6],SRK/T公式计算的为2