[Determination of procalcitonin (PCT) in micro-method on AQT90 FLEX® analyzer and in vitro stability study].

作者信息

Mine Louis, Petit Morgane, Nivet-Antoine Valérie, Beaudeux Jean-Louis, Hennequin Carole

机构信息

Laboratoire de biochimie générale, Hôpital universitaire Necker-Enfants Malades, AP-HP, Paris, France.

出版信息

Ann Biol Clin (Paris). 2018 Jan 1;76(1):53-59. doi: 10.1684/abc.2017.1308.

Abstract

Radiometer® has developed a point-of-care test for fast PCT measurement on whole blood in micromethod on a AQT90 FLEX® instrument. We have verified the analytical performances of the AQT90 FLEX® PCT assay in heparinized macrotube and EDTA microtube for pediatric use, according to modified French Society of Clinical Biology (SFBC) protocol to the requirements of the standard NF EN ISO 15189: 2012. The samples (n=61, 30 macrotubes, 31 microtubes) were analyzed by the Brahms Kryptor Compact Plus® reference method vs the AQT90 FLEX®. In a second step, we studied the stability of the PCT at room temperature for 24 h. A good correlation between the two methods on macro- or microtubes is observed (respectively r=0.990 and 0.993). The Bland-Altmann representations confirm the excellent correlation with a deviant, above the acceptable limit, which was calculated according to ISO 5725-6, for each type of tube and for the two concentration ranges (lower and greater than 1 ng/mL). The biases observed do not affect the clinical decision. No degradation of PCT after 24 h was demonstrated by the Mann and Whitney test on macro- and microtubes (p=0.50 and 0.34, respectively). The determination of PCT on AQT90 FLEX® has satisfactory analytical characteristics and can be used as point-of-care testing device on whole blood without pre-analytical treatment. In heparinized macrotube and EDTA microtube, the PCT is stable at room temperature up to 24 h.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验