Determining procalcitonin at point-of-care; A method comparison study of four commercial PCT assays.

作者信息

Li Chunbao, Huang Yaping, Xu Yubin

机构信息

KingMed Diagnostics Laboratory, Kangxin Road, Pudong New Area, Shanghai, China.

出版信息

Pract Lab Med. 2021 Mar 26;25:e00214. doi: 10.1016/j.plabm.2021.e00214. eCollection 2021 May.

Abstract

OBJECTIVE

Procalcitonin (PCT) testing adds value in the early detection of infection and sepsis, as well as in management of antibiotic therapy. We determined the analytical and diagnostic performance of four PCT assays at POC.

METHODS

PCT assays on AQT90 FLEX, Getein 1100, mLabs, and Finecare POC analyzers, in whole blood and plasma, were analyzed for repeatability, linearity, accuracy and concordance, by comparing with our reference PCT assay on the Cobas E602 system.

RESULTS

For all assays precision was found higher in plasma than in whole blood. AQT90 showed good performance in all analytical and diagnostic areas, irrespective of test matrix and PCT concentration. The other POC assays demonstrated at least one analytical weakness. The Getein assay showed adequate precision only in plasma at high PCT levels, the mLabs assay only in plasma at low PCT levels. Accuracy, as demonstrated by Bland-Altman and Passing-Bablok analysis, was found adequate only for the AQT90 FLEX and Getein 1100 assay. Diagnostic concordance at 0.5 ​ng/mL was found excellent for the AQT90 FLEX, Getein 1100, and Finecare assays, much lower for the mLabs assay. At 0.25 ​ng/mL, only the AQT90 FLEX and Finecare assays showed excellent concordance with the reference assay.

CONCLUSIONS

The AQT90 FLEX PCT assay demonstrated excellent analytical performance and diagnostic agreement with the Cobas E602 assay, allowing both stand-alone and side-by-side testing. The other assays demonstrated some analytical deficiencies, potentially limiting their diagnostic use. Any clinical use of PCT results should always be in combination with all other clinical signs and diagnostic information.

摘要

目的

降钙素原(PCT)检测在感染和脓毒症的早期检测以及抗生素治疗管理中具有重要价值。我们测定了四种即时检测(POC)的PCT检测方法的分析性能和诊断性能。

方法

通过与我们在Cobas E602系统上的参考PCT检测方法进行比较,分析了在AQT90 FLEX、Getein 1100、mLabs和Finecare POC分析仪上对全血和血浆进行的PCT检测的重复性、线性、准确性和一致性。

结果

所有检测方法在血浆中的精密度均高于全血。AQT90在所有分析和诊断领域均表现良好,与检测基质和PCT浓度无关。其他POC检测方法至少存在一个分析弱点。Getein检测方法仅在高PCT水平的血浆中具有足够的精密度,mLabs检测方法仅在低PCT水平的血浆中具有足够的精密度。通过Bland-Altman分析和Passing-Bablok分析证明,仅AQT90 FLEX和Getein 1100检测方法的准确性足够。在0.5 ng/mL时,AQT90 FLEX、Getein 1100和Finecare检测方法的诊断一致性极佳,mLabs检测方法的一致性则低得多。在0.25 ng/mL时,只有AQT90 FLEX和Finecare检测方法与参考检测方法具有极佳的一致性。

结论

AQT90 FLEX PCT检测方法与Cobas E602检测方法相比,具有出色的分析性能和诊断一致性,既可以单独检测,也可以并行检测。其他检测方法存在一些分析缺陷,可能会限制其诊断用途。PCT检测结果的任何临床应用都应始终结合所有其他临床体征和诊断信息。