Peking Union Medical College Hospital, Beijing, China.
Department of Pharmacology, Jinling Hospital, Medical School of Nanjing University, Nanjing, China.
Ann Lab Med. 2018 Mar;38(2):85-94. doi: 10.3343/alm.2018.38.2.85.
The immunosuppressant drugs (ISDs), tacrolimus and cyclosporine, are vital for solid organ transplant patients to prevent rejection. However, toxicity is a concern, and absorption is highly variable across patients; therefore, ISD levels need to be precisely monitored. In the Asia-Pacific (APAC) region, tacrolimus and cyclosporine concentrations are typically measured using immunoassays. The objective of this study was to assess the analytical performance of Roche Elecsystacrolimus and cyclosporinee electrochemiluminescence immunoassays (ECLIAs).
This evaluation was performed in seven centers across China, South Korea, and Malaysia. Imprecision (repeatability and reproducibility), assay accuracy, and lot-to-lot reagent variability were tested. The Elecsys ECLIAs were compared with commercially available immunoassays (Architect, Dimension, and Viva-E systems) using whole blood samples from patients with various transplant types (kidney, liver, heart, and bone marrow).
Coefficients of variation for repeatability and reproducibility were ≤5.4% and ≤12.4%, respectively, for the tacrolimus ECLIA, and ≤5.1% and ≤7.3%, respectively, for the cyclosporine ECLIA. Method comparisons of the tacrolimus ECLIA with Architect, Dimension, and Viva-E systems yielded slope values of 1.01, 1.14, and 0.897, respectively. The cyclosporine ECLIA showed even closer agreements with the Architect, Dimension, and Viva-E systems (slope values of 1.04, 1.04, and 1.09, respectively). No major differences were observed among the different transplant types.
The tacrolimus and cyclosporine ECLIAs demonstrated excellent precision and close agreement with other immunoassays tested. These results show that both assays are suitable for ISD monitoring in an APAC population across a range of different transplant types.
免疫抑制剂(ISD)药物,如他克莫司和环孢素,对于预防实体器官移植患者的排斥反应至关重要。然而,这些药物具有毒性,并且在患者之间的吸收情况存在很大差异,因此需要精确监测 ISD 水平。在亚太地区(APAC),通常使用免疫测定法来测量他克莫司和环孢素的浓度。本研究的目的是评估罗氏 Elecsys 他克莫司和环孢素电化学发光免疫分析(ECLIA)的分析性能。
本评估在中国、韩国和马来西亚的七个中心进行。对精密度(重复性和再现性)、测定准确性和批间试剂变异性进行了测试。Elecsys ECLIA 与来自各种移植类型(肾、肝、心和骨髓)患者的全血样本的市售免疫测定法(Architect、Dimension 和 Viva-E 系统)进行了比较。
他克莫司 ECLIA 的重复性和再现性的变异系数分别为≤5.4%和≤12.4%,环孢素 ECLIA 的分别为≤5.1%和≤7.3%。他克莫司 ECLIA 与 Architect、Dimension 和 Viva-E 系统的方法比较产生的斜率值分别为 1.01、1.14 和 0.897。环孢素 ECLIA 与 Architect、Dimension 和 Viva-E 系统的一致性更好(斜率值分别为 1.04、1.04 和 1.09)。不同移植类型之间没有观察到明显差异。
他克莫司和环孢素 ECLIAs 表现出优异的精密度,与测试的其他免疫测定法具有密切的一致性。这些结果表明,两种测定法均适用于亚太地区不同移植类型患者的 ISD 监测。