Abstract

OBJECTIVE

To assess the validity and reliability of Doppler ultrasonography (DOP) as compared with invasive arterial blood pressure measurements in anaesthetized dogs weighing less than 5 kg.

STUDY DESIGN

Controlled, prospective, clinical study.

ANIMALS

A total of 41 privately owned dogs weighing less than 5 kg.

METHODS

The dogs were anaesthetized, and an intra-arterial catheter was placed aseptically in the dorsal pedal artery of the pelvic limb to perform invasive blood pressure (IBP) measurement. The contralateral metatarsal surface of the foot was clipped in order to perform DOP. Both techniques were used to record blood pressure measurements every 5 minutes during surgical procedures. The blood pressure measurements were categorized into two groups: hypotensive [mean arterial pressure (MAP) < 60 mmHg] and normotensive (MAP between 60 and 120 mmHg). A linear mixed model was used to compare the DOP and IBP values. The results were evaluated according to the requirements of the American College of Veterinary Internal Medicine (ACVIM) guidelines for the validation of devices.

RESULTS

DOP provided higher values compared to the systolic arterial blood pressure (SAP) and MAP of IBP measurements. The closest agreement between the two techniques was found for SAP; the bias was 8.8, and limits of agreement (LOA) were -32.9 and 50.4. Similar results were observed when the IBP technique was categorized. The closest agreement was for SAP in animals categorized as normotensive; the bias was 8.2, and LOAs were -32.8 and 49.2. The level of agreement between DOP and IBP did not meet the ACVIM recommendations.

CONCLUSIONS AND CLINICAL RELEVANCE

Our results suggest there is poor agreement between DOP and IBP measurements in anaesthetized dogs weighing less than 5 kg. Hence, the use of DOP in these animals could be misleading.

摘要

目的

评估在体重小于5kg的麻醉犬中,与有创动脉血压测量相比,多普勒超声检查(DOP)的有效性和可靠性。

研究设计

对照、前瞻性临床研究。

动物

总共41只体重小于5kg的私人饲养犬。

方法

对犬进行麻醉,无菌条件下在盆腔肢的足背动脉放置动脉内导管以进行有创血压(IBP)测量。修剪对侧足的跖面以进行DOP检查。在手术过程中,每隔5分钟使用这两种技术记录血压测量值。血压测量值分为两组:低血压组[平均动脉压(MAP)<60mmHg]和正常血压组(MAP在60至120mmHg之间)。使用线性混合模型比较DOP和IBP值。根据美国兽医内科学会(ACVIM)设备验证指南的要求评估结果。

结果

与IBP测量的收缩期动脉血压(SAP)和MAP相比,DOP提供的值更高。两种技术之间在SAP方面的一致性最为接近;偏差为8.8,一致性界限(LOA)为-32.9和50.4。当对IBP技术进行分类时,观察到类似结果。在分类为正常血压的动物中,SAP的一致性最为接近;偏差为8.2,LOA为-32.8和49.2。DOP和IBP之间的一致性水平未达到ACVIM的建议。

结论及临床意义

我们的结果表明,在体重小于5kg的麻醉犬中,DOP和IBP测量之间的一致性较差。因此,在这些动物中使用DOP可能会产生误导。