Avoidance from palatal blood vessel rupture is a major concern during the palatal soft tissue graft surgery. There is no defined chair-side and case-specific palatal blood vessel detection approach to facilitate the harvesting process. The objective of this pilot study is to assess the feasibility of a near-infrared vein visualization system in the screening process of palatal blood vessels.
An extraoral vein visualization device (AccuVein AV400) was applied to a total of 304 hemi-maxilla of 152 individuals by two blind examiners. The study groups were classified according to their maximum inter-incisal measurements. The distances between the coronal border of the vessel image and the mid-palatal gingival margins of the adjacent teeth were measured and in each group. The correlations among the measurements were evaluated within groups.
The blood vessel to the adjacent teeth measurements exhibited no statistical difference between both examiners in all subjects (p<0.001). Correlations between the examiners gradually increased in all groups as the mouth opening rates of the subjects were increased (p<0.001).
In the current state, screening of the palatal blood vessels via near-infrared vein visualization technology seems to be not suitable for every individual due to the restrictive effect of mouth opening. However, the promising results of this preliminary study demonstrated increasing consistency between the measurements of the examiners as the inter-incisal distance increase which emphasized the need an intraoral version of the device. Considering the lack of local decision-making technology for the detection of palatal blood vessels, further studies are required for development and optimization of these systems.
避免腭部血管破裂是腭部软组织移植手术中的一个主要问题。目前尚无明确的椅旁且针对具体病例的腭部血管检测方法来辅助采集过程。本初步研究的目的是评估近红外静脉可视化系统在腭部血管筛查过程中的可行性。
两名盲法检查者将一种口外静脉可视化设备(AccuVein AV400)应用于152名个体的总共304个半上颌骨。研究组根据其最大切牙间测量值进行分类。测量每组中血管图像的冠状边界与相邻牙齿的腭中牙龈边缘之间的距离。评估组内测量值之间的相关性。
在所有受试者中,两位检查者对相邻牙齿的血管测量值均无统计学差异(p<0.001)。随着受试者张口率的增加,所有组中检查者之间的相关性逐渐增加(p<0.001)。
在当前状态下,由于张口的限制作用,通过近红外静脉可视化技术筛查腭部血管似乎并不适用于每个个体。然而,这项初步研究的良好结果表明,随着切牙间距离增加,检查者测量值之间的一致性增强,这强调了需要该设备的口内版本。鉴于缺乏用于检测腭部血管的局部决策技术,需要进一步研究来开发和优化这些系统。