血糖自我监测改进系统的系统准确性和用户性能评估

System Accuracy and User Performance Evaluation of an Improved System for Self-Monitoring of Blood Glucose.

作者信息

Baumstark Annette, Jendrike Nina, Liebing Christina, Haug Cornelia, Freckmann Guido

机构信息

1 Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.

出版信息

J Diabetes Sci Technol. 2018 Mar;12(2):407-411. doi: 10.1177/1932296817752958. Epub 2018 Jan 1.

Abstract

An improved test cassette for the integrated Accu-Chek® Mobile system (Roche Diabetes Care GmbH, Mannheim, Germany) has been developed. System accuracy of this improved system was evaluated based on ISO 15197:2013, clause 6.3, for three reagent system lots. According to this standard, at least 95% of the system's measurement results shall be within ±15 mg/dL and ±15% of the results of the comparison method at glucose concentrations <100 mg/dL and ≥100 mg/dL (accuracy criterion A), respectively, and at least 99% of results shall be within consensus error grid zones A and B (accuracy criterion B). In addition, accuracy was evaluated in the hands of users based on ISO 15197:2013, clause 8, with one reagent system lot.

摘要

已开发出一种用于集成式Accu-Chek® Mobile系统(德国曼海姆罗氏糖尿病护理有限公司)的改进型测试盒。基于ISO 15197:2013第6.3条,对三个试剂系统批次评估了这种改进系统的系统准确性。根据该标准,在葡萄糖浓度<100 mg/dL和≥100 mg/dL时,系统测量结果中至少95%应分别在比较方法结果的±15 mg/dL和±15%范围内(准确性标准A),并且至少99%的结果应在一致性误差网格区域A和B内(准确性标准B)。此外,基于ISO 15197:2013第8条,使用一个试剂系统批次在用户手中评估了准确性。

相似文献

1
System Accuracy and User Performance Evaluation of an Improved System for Self-Monitoring of Blood Glucose.
J Diabetes Sci Technol. 2018 Mar;12(2):407-411. doi: 10.1177/1932296817752958. Epub 2018 Jan 1.
3
System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197.
Diabetes Technol Ther. 2010 Mar;12(3):221-31. doi: 10.1089/dia.2009.0128.
6
Performance of two updated blood glucose monitoring systems: an evaluation following ISO 15197:2013.
Curr Med Res Opin. 2016 May;32(5):847-55. doi: 10.1185/03007995.2016.1146666. Epub 2016 Feb 23.
8
The accuracy evaluation of four blood glucose monitoring systems according to ISO 15197:2003 and ISO 15197:2013 criteria.
Prim Care Diabetes. 2019 Jun;13(3):252-258. doi: 10.1016/j.pcd.2018.12.010. Epub 2019 Feb 13.
9
Evaluation of 12 blood glucose monitoring systems for self-testing: system accuracy and measurement reproducibility.
Diabetes Technol Ther. 2014 Feb;16(2):113-22. doi: 10.1089/dia.2013.0208. Epub 2013 Nov 8.

引用本文的文献

2
Flash glucose monitoring system in gestational diabetes: a study of accuracy and usability.
Hormones (Athens). 2023 Dec;22(4):703-713. doi: 10.1007/s42000-023-00485-z. Epub 2023 Sep 23.
3
Clinical Study of a High Accuracy Green Design Blood Glucose Monitor Using an Innovative Optical Transmission Absorbance System.
J Diabetes Sci Technol. 2022 Sep;16(5):1069-1075. doi: 10.1177/19322968211060865. Epub 2021 Dec 10.

本文引用的文献

4
Performance of a new blood glucose monitoring system in the hands of intended users.
Diabetes Technol Ther. 2012 Sep;14(9):783-9. doi: 10.1089/dia.2012.0046. Epub 2012 Jun 29.
5
Influence of fruit juice on fingertips and patient behavior on self-monitoring of blood glucose.
Diabetes Res Clin Pract. 2012 May;96(2):e50-2. doi: 10.1016/j.diabres.2012.01.028. Epub 2012 Feb 18.
7
Factors affecting blood glucose monitoring: sources of errors in measurement.
J Diabetes Sci Technol. 2009 Jul 1;3(4):903-13. doi: 10.1177/193229680900300438.
8
Standardized evaluation of nine instruments for self-monitoring of blood glucose.
Diabetes Technol Ther. 2008 Dec;10(6):467-77. doi: 10.1089/dia.2008.0034.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验