Ryan James D, Tempel Nathan D, Engle Dean D, Hayes David L, Cha Yong-Mei, Asirvatham Samuel J
Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN, USA.
Pacing Clin Electrophysiol. 2018 Feb;41(2):161-166. doi: 10.1111/pace.13269. Epub 2018 Jan 24.
Transthoracic impedance measurements (TIM) is primarily used for minute ventilation rate adaptive sensors in pacemakers. With elevated impedance, the TIM electrical signal itself is oversensed, causing device malfunction.
We report an increased incidence of TIM oversensing.
Retrospective chart review. We review existing records of 18 patients who have demonstrated device malfunction with TIM oversensing.
We have found a 1.8% incidence of TIM-related oversensing in our patient population of 959 patients with contemporary Boston Scientific (Marlborough, MA, USA) pacemakers and cardiac resynchronization therapy pacemakers. One patient experienced a syncopal episode.
Oversensing with pacing inhibition is apparent with the potential of adverse effects to patients.