Potential effects of low-dose average CT on cardiac implantable electronic devices.

机构信息

Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Unit 1352, Houston, TX, 70030-4009, USA.

Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, 70030, USA.

出版信息

J Nucl Cardiol. 2019 Aug;26(4):1161-1165. doi: 10.1007/s12350-018-1186-y. Epub 2018 Jan 17.

Abstract

BACKGROUND

Average CT has been shown to be more accurate than conventional helical CT in quantitation of the PET data. The risk of CT irradiation of a cardiac implantable electronic device (CIED) causing an adverse event is low and is generally outweighed by the clinical benefit of a medically indicated examination. However, irradiation of CIED over one breath cycle in cine CT scan for average CT could impose risks on a patient who is pacing dependent. The purpose of this study was to demonstrate that low-dose average CT can be safe for CIED.

METHODS

A Medtronic CIED of model Protecta VR was submerged in a saline bath for a series of 4-s cine CT scans on a GE CT scanner programmed to deliver a 2-cm-wide radiation at a dose rate of 0.9 to 41.2 mGy/s to the CIED. The number of over-sensings was recorded as the interference of radiation to the CIED.

RESULTS

Dose rates ≥ 1.9 mGy/s caused over-sensing. The higher the dose rate, the more over-sensings there were. The lowest dose rate of 0.9 mGy/s did not cause any over-sensing.

CONCLUSIONS

Low-dose average CT at 0.9 mGy/s can be safe for a CIED patient who is pacing dependent.

摘要

背景

平均 CT 已被证明比传统螺旋 CT 更能准确量化 PET 数据。心脏植入式电子设备(CIED)受到 CT 辐射而发生不良事件的风险较低,且通常低于医学检查的临床获益。然而,在电影 CT 扫描中,对 CIED 进行一个呼吸周期的平均 CT 照射可能会对依赖起搏的患者造成风险。本研究旨在证明低剂量平均 CT 对 CIED 是安全的。

方法

将 Medtronic 的 CIED Protecta VR 模型浸泡在盐水中,在 GE CT 扫描仪上进行一系列 4 秒电影 CT 扫描,该 CT 扫描仪编程为以 0.9 至 41.2 mGy/s 的剂量率向 CIED 输送 2cm 宽的辐射。记录超感次数作为辐射对 CIED 的干扰。

结果

剂量率≥1.9 mGy/s 会引起超感。剂量率越高,超感次数越多。最低剂量率 0.9 mGy/s 不会引起任何超感。

结论

对于依赖起搏的 CIED 患者,0.9 mGy/s 的低剂量平均 CT 是安全的。