评估环境光对口腔内扫描仪扫描准确性、精度和时间的影响。

Evaluating the influence of ambient light on scanning trueness, precision, and time of intra oral scanner.

机构信息

Gerodontology and Oral Rehabilitation, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Tokyo, Japan.

Gerodontology and Oral Rehabilitation, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Tokyo, Japan.

出版信息

J Prosthodont Res. 2018 Jul;62(3):324-329. doi: 10.1016/j.jpor.2017.12.005. Epub 2018 Feb 1.

Abstract

PURPOSE

This study evaluated the influence of illuminance and color temperature of ambient light on the trueness, precision, and scanning time of a digital impression.

METHODS

Master data were acquired with a high-accuracy coordinate-measuring machine. The illuminance of ambient light was set at 0lux, 500lux, and 2500lux with a light-emitting diode (LED). Using a conversion filter, the color temperature was set at 3900 Kelvin (K) (yellow), 4100K (orange), 7500K (white), and 19,000K (blue). There were thus a total of 12 possible lighting conditions. The reference model was scanned five times under each condition by an intraoral scanner. Trueness was calculated as the mean difference between the master data and experimental data. Precision was calculated as the mean difference between the repeated scans in each test group. Statistical analysis was performed with two-way analysis of variance (ANOVA) and post hoc Tukey's multiple comparison test. The significance level was 0.05.

RESULTS

For trueness, the mean deviation was significantly lower at 500lux than at 0lux and 2500lux. At 500lux, the mean deviation was significantly lower at 3900K than at other temperatures. Regardless of the color temperature, the scanning time was significantly longer at 2500lux than at other illuminance levels.

CONCLUSIONS

The 3900K and 500lux condition is the most appropriate lighting condition for taking a digital impression. This condition is typical of clinical settings. High illuminance ambient light increased the scanning time.

摘要

目的

本研究评估环境光照度和色温对数字化印模准确性、精密度和扫描时间的影响。

方法

采用高精度三坐标测量机获取主数据。采用发光二极管(LED)将环境光照度设置为 0lux、500lux 和 2500lux,通过转换滤光片将色温设置为 3900 开尔文(K)(黄色)、4100K(橙色)、7500K(白色)和 19000K(蓝色)。因此,共有 12 种可能的照明条件。每个条件下通过口腔内扫描仪对参考模型进行五次扫描。准确性计算为主数据和实验数据之间的平均差值。精密度计算为每组重复扫描之间的平均差值。采用双因素方差分析(ANOVA)和事后 Tukey 多重比较检验进行统计分析。显著性水平为 0.05。

结果

对于准确性,500lux 时的平均偏差明显低于 0lux 和 2500lux。在 500lux 时,3900K 时的平均偏差明显低于其他温度。无论色温如何,2500lux 时的扫描时间明显长于其他光照水平。

结论

3900K 和 500lux 的条件是获取数字化印模的最佳照明条件。这种条件是临床环境中的典型条件。高环境光照度增加了扫描时间。