To determine the influence of different scanning and ambient light conditions on the trueness values of an intraoral scanning (IOS) device.
The present study was conducted among 20 complete dentate volunteers. After making complete maxillary arch impressions with vinyl polysiloxane material, type IV dental stone was poured, and working casts were obtained. Then, the models were digitized using a dental laboratory scanner (LS) and standard tessellation language (STL) files were acquired. Full arch digital scans of the volunteers were also performed using an IOS device with 2 ambient light conditions (RL: room light, 1003 lux, and ZL: zero light, 0 lux) and 2 scanning light modes (blue [B] and white [W]). Twenty digital scans per group at each light condition were consecutively obtained. Discrepancy between the digital scans and digitized working casts was analyzed using a 3D inspection software. The obtained data were analyzed using one-way ANOVA and Tukey tests (α = 0.05).
Significant differences were obtained among different light conditions for the IOS device (p ˂ 0.0001, F = 42.958 for positive deviations and F = 29.278, for negative deviations). The room light and blue mode (RLB) condition had the lowest deviation values compared with the other light conditions, followed by room light and white mode (RLW), zero light and blue mode (ZLB), and zero light and white mode (ZLW; p = 0.008, p ˂ 0.0001, and p ˂ 0.0001, respectively).
Lighting conditions are effective in assessing the trueness of the digital impressions made with an IOS device. RLB conditions are recommended for taking a digital impression.
确定不同扫描和环境光条件对口腔内扫描(IOS)设备准确性值的影响。
本研究在 20 名完全无牙颌志愿者中进行。使用乙烯基聚硅氧烷材料制取上颌全牙弓印模后,倒入 IV 型牙科石膏,获得工作模型。然后使用牙科实验室扫描仪(LS)对模型进行数字化,并获取标准网格语言(STL)文件。还使用具有 2 种环境光条件(RL:室内光,1003 勒克斯,ZL:零光,0 勒克斯)和 2 种扫描光模式(蓝色[B]和白色[W])的 IOS 设备对志愿者进行全牙弓数字扫描。在每种光照条件下,每个组连续获得 20 个数字扫描。使用三维检测软件分析数字扫描与数字化工作模型之间的差异。使用单向方差分析和 Tukey 检验(α=0.05)对获得的数据进行分析。
IOS 设备在不同光照条件下获得了显著差异(p ˂0.0001,正偏差的 F = 42.958,负偏差的 F = 29.278)。与其他光照条件相比,室内光和蓝色模式(RLB)条件下的偏差值最低,其次是室内光和白色模式(RLW)、零光和蓝色模式(ZLB)以及零光和白色模式(ZLW;p = 0.008,p ˂0.0001,p ˂0.0001)。
光照条件会影响 IOS 设备制取的数字印模的准确性。建议在采用数字印模时使用 RLB 条件。