No Difference in Conventional Polyethylene Wear Between Yttria-stabilized Zirconia and Cobalt-chromium-molybdenum Femoral Heads at 10 Years.

作者信息

Morrison Todd A, Moore Rebecca D, Meng Jia, Rimnac Clare M, Kraay Matthew J

机构信息

1Department of Orthopaedic Surgery, University Hospitals Case Medical Center and Case Western Reserve University, 11100 Euclid Avenue, Cleveland, OH 44106 USA.

2Case Western Reserve University School of Medicine, Cleveland, OH USA.

出版信息

HSS J. 2018 Feb;14(1):60-66. doi: 10.1007/s11420-017-9579-z. Epub 2017 Oct 13.

Abstract

BACKGROUND

Concerns have arisen regarding deterioration of wear properties of yttria-stabilized zirconia (YSZ) femoral head on conventional polyethylene (PE) bearings due to YSZ phase transformation.

QUESTIONS/PURPOSES: The purpose of this study was to determine if there is a difference in long-term PE wear properties between YSZ and cobalt-chromium-molybdenum (Co-Cr-Mo) femoral heads.

METHODS

Ten-year radiographic wear assessment was performed on a cohort of patients enrolled in a prospective randomized clinical trial comparing total hip arthroplasty with YSZ or Co-Cr-Mo femoral heads on conventional, non-cross-linked PE.

RESULTS

PE linear wear, annualized wear, and steady-state wear rates remained low and similar between groups. No cases of osteolysis were observed.

CONCLUSIONS

Measured conventional PE wear was similar between YSZ and Co-Cr-Mo femoral heads with the steady-state wear rates for both remaining below the generally accepted threshold at which osteolysis typically occurs. Whether clinically relevant phase transformation with YSZ femoral heads occurs is uncertain; however, the use of YSZ femoral heads in this study was not associated with increased PE wear, osteolysis, or deterioration of wear properties.

摘要

背景

由于氧化钇稳定氧化锆(YSZ)的相变,人们对传统聚乙烯(PE)轴承上的 YSZ 股骨头磨损性能的恶化产生了担忧。

问题/目的:本研究的目的是确定 YSZ 股骨头和钴铬钼(Co-Cr-Mo)股骨头在长期 PE 磨损性能上是否存在差异。

方法

对一组参加前瞻性随机临床试验的患者进行了为期十年的影像学磨损评估,该试验比较了在传统非交联 PE 上使用 YSZ 或 Co-Cr-Mo 股骨头的全髋关节置换术。

结果

两组之间的 PE 线性磨损、年化磨损和稳态磨损率均保持在较低水平且相似。未观察到骨溶解病例。

结论

YSZ 和 Co-Cr-Mo 股骨头之间测得的传统 PE 磨损相似,两者的稳态磨损率均低于通常公认的骨溶解发生阈值。YSZ 股骨头是否会发生临床相关的相变尚不确定;然而,在本研究中使用 YSZ 股骨头与 PE 磨损增加、骨溶解或磨损性能恶化无关。