Aşangil Derya, Hüdai Taşdemir İbrahim, Kılıç Esma
Ankara University, Faculty of Science, Department of Chemistry, 06100 Ankara, Turkey.
Amasya University, Faculty of Arts and Science, Department of Chemistry, 05100 Amasya, Turkey.
J Pharm Anal. 2012 Jun;2(3):193-199. doi: 10.1016/j.jpha.2012.01.009. Epub 2012 Feb 8.
Anodic behavior of aripiprazole (ARP) was studied using electrochemical methods. Charge transfer, diffusion and surface coverage coefficients of adsorbed molecules and the number of electrons transferred in electrode mechanisms were calculated for quasi-reversible and adsorption-controlled electrochemical oxidation of ARP at 1.15 V versus Ag/AgCl at pH 4.0 in Britton-Robinson buffer (BR) on glassy carbon electrode. Voltammetric methods for direct determination of ARP in pharmaceutical dosage forms and biological samples were developed. Linearity range is found as from 11.4 μM (5.11 mg/L) to 157 μM (70.41 mg/L) without stripping mode and it is found as from 0.221 μM (0.10 mg/L) to 13.6 μM (6.10 mg/L) with stripping mode. Limit of detection (LOD) was found to be 0.11 μM (0.05 mg/L) in stripping voltammetry. Methods were successfully applied to assay the drug in tablets, human serum and human urine with good recoveries between 95.0% and 104.6% with relative standard deviation less than 10%.
采用电化学方法研究了阿立哌唑(ARP)的阳极行为。计算了在pH 4.0的Britton-Robinson缓冲液(BR)中,玻碳电极上相对于Ag/AgCl在1.15 V时ARP的准可逆和吸附控制电化学氧化过程中吸附分子的电荷转移、扩散和表面覆盖系数以及电极反应中转移的电子数。建立了直接测定药物剂型和生物样品中ARP的伏安法。在无溶出模式下,线性范围为11.4 μM(5.11 mg/L)至157 μM(70.41 mg/L);在有溶出模式下,线性范围为0.221 μM(0.10 mg/L)至13.6 μM(6.10 mg/L)。溶出伏安法的检测限(LOD)为0.11 μM(0.05 mg/L)。该方法成功应用于片剂、人血清和人尿中药物的测定,回收率良好,在95.0%至104.6%之间,相对标准偏差小于10%。