To report the 3-year assessment of feasibility and usefulness of microscope-integrated intraoperative OCT (iOCT) during ophthalmic surgery.
Prospective, consecutive case series.
Adult participants undergoing incisional ophthalmic surgery with iOCT imaging who consented to be enrolled in the Determination of Feasibility of Intraoperative Spectral-Domain Microscope Combined/Integrated OCT Visualization during En Face Retinal and Ophthalmic Surgery (DISCOVER) study.
The DISCOVER study is a single-site, multisurgeon, institutional review board-approved investigational device prospective study. Participants included patients undergoing anterior or posterior segment surgery who underwent iOCT imaging with 1 of 3 prototype microscope-integrated iOCT systems (i.e., Zeiss Rescan 700, Leica EnFocus, or Cole Eye iOCT systems). Clinical characteristics were documented, iOCT was directed by the operating surgeon at predetermined surgical time points, and each surgeon completed a questionnaire after surgery to evaluate the usefulness of iOCT during surgery.
Feasibility of iOCT based ability to obtain an OCT image during surgery and usefulness of iOCT based on surgeon reporting during surgery.
Eight hundred thirty-seven eyes (244 anterior segment cases and 593 posterior segment cases) were enrolled in the DISCOVER study. Intraoperative OCT demonstrated feasibility with successful image acquisition in 820 eyes (98.0%; 95% confidence interval [CI], 96.8%-98.8%). In 106 anterior segment cases (43.4%; 95% CI, 37.1%-49.9%), the surgeons indicated that the iOCT information impacted their surgical decision making and altered the procedure. In posterior segment procedures, surgeons reported that iOCT enabled altered surgical decision making during the procedure in 173 cases (29.2%; 95% CI, 25.5%-33.0%).
The DISCOVER iOCT study demonstrated both generalized feasibility and usefulness based on the surgeon-reported impact on surgical decision making. This large-scale study confirmed similar findings from other studies on the potential value and impact of iOCT on ophthalmic surgery.
报告显微镜集成术中光学相干断层扫描(iOCT)在眼科手术中 3 年的可行性和实用性评估。
前瞻性、连续病例系列。
同意参加确定术中光谱域显微镜联合/集成 OCT 可视化在面向视网膜和眼科手术中的可行性(DISCOVER)研究的接受切口眼科手术的成年参与者,并进行 iOCT 成像。
DISCOVER 研究是一项单站点、多外科医生、机构审查委员会批准的研究器械前瞻性研究。参与者包括接受前节或后节手术的患者,他们使用 3 种原型显微镜集成 iOCT 系统中的 1 种进行 iOCT 成像(即蔡司 Rescan 700、徕卡 EnFocus 或 Cole Eye iOCT 系统)。记录临床特征,由手术医生在预定的手术时间点指导 iOCT,每位医生在手术后完成一份问卷,以评估 iOCT 在手术中的实用性。
基于术中获得 OCT 图像的 iOCT 能力的可行性,以及基于手术医生术中报告的 iOCT 的有用性。
共纳入 837 只眼(244 例前节病例和 593 例后节病例)参加 DISCOVER 研究。术中 OCT 显示出可行性,820 只眼(98.0%;95%置信区间[CI],96.8%-98.8%)成功获得图像。在前节 106 例病例中(43.4%;95%CI,37.1%-49.9%),外科医生表示 iOCT 信息影响了他们的手术决策并改变了手术过程。在后节手术中,外科医生报告在 173 例病例(29.2%;95%CI,25.5%-33.0%)中,iOCT 使手术过程中的手术决策发生改变。
基于外科医生报告对手术决策的影响,DISCOVER iOCT 研究证明了其具有广泛的可行性和实用性。这项大规模研究证实了其他关于 iOCT 对眼科手术潜在价值和影响的研究的类似发现。