Hofstetter Lorne W, Yeo Desmond T B, Dixon W Thomas, Marinelli Luca, Foo Thomas K
Department of Radiology, University of Utah, Salt Lake City, UT, USA.
GE Global Research, Niskayuna, NY, USA.
Magn Reson Imaging. 2018 Jun;49:86-93. doi: 10.1016/j.mri.2018.01.002. Epub 2018 Feb 2.
A three-point image reconstruction method for internally referenced MR thermometry was developed. The technique exploits the fact that temperature-induced changes in the water resonance frequency are small relative to the chemical shift difference between water and fat signals. This property enabled the use of small angle approximations to derive an analytic phase-based fat-water separation method for MR thermometry. Ethylene glycol and cream cool-down experiments were performed to validate measurement technique. Over a cool-down temperature range of 20 °C, maximum deviation between probe and MR measurement (averaged over 1.3 cm region surrounding probe) was 0.6 °C and 1.1 °C for ethylene glycol and cream samples, respectively.