自动监测仪测量南美的儿科人群血压准确吗?SAYCARE 研究。

Is the Measurement of Blood Pressure by Automatic Monitor in the South American Pediatric Population Accurate? SAYCARE Study.

机构信息

DOMEN (Metabolic Diseases, Exercise and Nutrition) Research Group, Center for Health Sciences, Federal University of Piaui, Teresina, Brazil.

Youth/Child and Cardiovascular Risk and Environmental Research Group, School of Medicine, University of São Paulo, São Paulo, Brazil.

出版信息

Obesity (Silver Spring). 2018 Mar;26 Suppl 1:S41-S46. doi: 10.1002/oby.22119.

Abstract

OBJECTIVE

This study aimed to test the validity of an automatic oscillometric device to measure the blood pressure (BP) in children (n = 191) and adolescents (n = 127) aged 3 to 18 years.

METHODS

Systolic BP (SBP) and diastolic BP (DBP) levels were measured simultaneous by automatic device and mercury column with Y-connection. To verify the validity, Bland-Altman plots and limits of agreement of 95% (95% LOA), specificity and sensitivity of the device, and the grade of British Hypertension Society (BHS) criteria were used.

RESULTS

The monitor measurements demonstrated lower measurement bias (mean difference [95% LOA]): 1.4 (-9.9 to 12.8) mmHg in children and 4.3 (-7.8 to 16.5) mmHg in adolescents for SBP. For DBP, it was 2.2 (-7.4 to 11.7) mmHg in children and 1.4 (-8.4 to 11.1) mmHg in adolescents. The sensitivity in children was 21.4 (95% CI = 16.3-26.6), and in adolescents, it was 20.0 (95% CI = 13.2-26.8); the specificity was 95.9 (95% CI = 93.4-98.4) in children and 100.0 (95% CI = 100.0-100.0) in adolescents. The monitor-tested ratings are Grade B for SBP in children and SBP and DBP in adolescents and Grade C for DBP in children.

CONCLUSIONS

The automatic monitor presented high values of specificity and lower values of sensitivity to the diagnosis of HBP; however, it can be considered accurate (lower measurement bias) and valid for epidemiological and clinical practice in accordance with BHS criteria.

摘要

目的

本研究旨在检验一种自动测振仪测量 3 至 18 岁儿童(n=191)和青少年(n=127)血压(BP)的有效性。

方法

采用自动测振仪与 Y 型三通管水银柱同时测量收缩压(SBP)和舒张压(DBP)。为了验证其有效性,采用 Bland-Altman 图和 95%(95%LOA)一致性界限、设备的特异性和敏感性以及英国高血压学会(BHS)标准的等级来评估。

结果

该监测仪的测量值显示出较低的测量偏差(平均差值[95%LOA]):儿童为 1.4(-9.9 至 12.8)mmHg,青少年为 4.3(-7.8 至 16.5)mmHg,SBP。对于 DBP,儿童为 2.2(-7.4 至 11.7)mmHg,青少年为 1.4(-8.4 至 11.1)mmHg。儿童的敏感性为 21.4(95%CI=16.3-26.6),青少年为 20.0(95%CI=13.2-26.8);特异性在儿童中为 95.9(95%CI=93.4-98.4),在青少年中为 100.0(95%CI=100.0-100.0)。根据 BHS 标准,监测仪的测试结果为儿童 SBP 为 B 级,青少年 SBP 和 DBP 为 B 级,儿童 DBP 为 C 级。

结论

该自动监测仪对高血压的诊断具有较高的特异性和较低的敏感性,但根据 BHS 标准,其测量偏差较低,可用于流行病学和临床实践,具有准确性。