Obesity (Silver Spring). 2018 Mar;26 Suppl 1:S41-S46. doi: 10.1002/oby.22119.
This study aimed to test the validity of an automatic oscillometric device to measure the blood pressure (BP) in children (n = 191) and adolescents (n = 127) aged 3 to 18 years.
Systolic BP (SBP) and diastolic BP (DBP) levels were measured simultaneous by automatic device and mercury column with Y-connection. To verify the validity, Bland-Altman plots and limits of agreement of 95% (95% LOA), specificity and sensitivity of the device, and the grade of British Hypertension Society (BHS) criteria were used.
The monitor measurements demonstrated lower measurement bias (mean difference [95% LOA]): 1.4 (-9.9 to 12.8) mmHg in children and 4.3 (-7.8 to 16.5) mmHg in adolescents for SBP. For DBP, it was 2.2 (-7.4 to 11.7) mmHg in children and 1.4 (-8.4 to 11.1) mmHg in adolescents. The sensitivity in children was 21.4 (95% CI = 16.3-26.6), and in adolescents, it was 20.0 (95% CI = 13.2-26.8); the specificity was 95.9 (95% CI = 93.4-98.4) in children and 100.0 (95% CI = 100.0-100.0) in adolescents. The monitor-tested ratings are Grade B for SBP in children and SBP and DBP in adolescents and Grade C for DBP in children.
The automatic monitor presented high values of specificity and lower values of sensitivity to the diagnosis of HBP; however, it can be considered accurate (lower measurement bias) and valid for epidemiological and clinical practice in accordance with BHS criteria.
本研究旨在检验一种自动测振仪测量 3 至 18 岁儿童(n=191)和青少年(n=127)血压(BP)的有效性。
采用自动测振仪与 Y 型三通管水银柱同时测量收缩压(SBP)和舒张压(DBP)。为了验证其有效性,采用 Bland-Altman 图和 95%(95%LOA)一致性界限、设备的特异性和敏感性以及英国高血压学会(BHS)标准的等级来评估。
该监测仪的测量值显示出较低的测量偏差(平均差值[95%LOA]):儿童为 1.4(-9.9 至 12.8)mmHg,青少年为 4.3(-7.8 至 16.5)mmHg,SBP。对于 DBP,儿童为 2.2(-7.4 至 11.7)mmHg,青少年为 1.4(-8.4 至 11.1)mmHg。儿童的敏感性为 21.4(95%CI=16.3-26.6),青少年为 20.0(95%CI=13.2-26.8);特异性在儿童中为 95.9(95%CI=93.4-98.4),在青少年中为 100.0(95%CI=100.0-100.0)。根据 BHS 标准,监测仪的测试结果为儿童 SBP 为 B 级,青少年 SBP 和 DBP 为 B 级,儿童 DBP 为 C 级。
该自动监测仪对高血压的诊断具有较高的特异性和较低的敏感性,但根据 BHS 标准,其测量偏差较低,可用于流行病学和临床实践,具有准确性。