Evaluation of a Portable Doppler Ultrasound Gating Device for Fetal Cardiac MR Imaging: Initial Results at 1.5T and 3T.

机构信息

Department of Diagnostic and Interventional Radiology and Nuclear Medicine, University Medical Center Hamburg-Eppendorf.

northh medical GmbH.

出版信息

Magn Reson Med Sci. 2018 Oct 10;17(4):308-317. doi: 10.2463/mrms.mp.2017-0100. Epub 2018 Feb 21.

Abstract

PURPOSE

Fetal cardiac MRI has the potential to play an important role in the assessment of fetal cardiac pathologies, but it is up to now not feasible due to a missing gating method. The purpose of this work was the evaluation of Doppler ultrasound (DUS) for external fetal cardiac gating with regard to compatibility, functionality, and reliability. Preliminary results were assessed performing fetal cardiac MRI.

METHODS

An MRI conditional DUS device was developed to obtain a gating signal from the fetal heart. The MRI compatibility was evaluated at 1.5T and 3T using B field maps and gradient echo images. The quality and sensitivity of the DUS device to detect the fetal heart motion for cardiac gating were evaluated outside the MRI room in 15 fetuses. A dynamic fetal cardiac phantom was employed to evaluate distortions of the DUS device and gating signal due to electromagnetic interferences at 1.5T and 3T. In the first in vivo experience, dynamic fetal cardiac images were acquired in four-chamber view at 1.5T and 3T in two fetuses.

RESULTS

The maximum change in the B field and signal intensity with and without the DUS device was <6.5% for 1.5T and 3T. The sensitivity of the DUS device to detect the fetal heartbeat was 99.1%. Validation of the DUS device using the fetal cardiac phantom revealed no electromagnetic interferences at 1.5T or 3T and a high correlation to the simulated heart frequencies. Fetal cardiac cine images were successfully applied and showed good image quality.

CONCLUSION

An MR conditional DUS gating device was developed and evaluated revealing safety, compatibility, and reliability for different field strengths. In a preliminary experience, the DUS device was successfully applied for in vivo fetal cardiac imaging at 1.5T and 3T.

摘要

目的

胎儿心脏 MRI 有可能在评估胎儿心脏病变方面发挥重要作用,但由于缺乏门控方法,目前还不可行。本研究旨在评估多普勒超声(DUS)在外源性胎儿心脏门控中的适用性、功能性和可靠性。初步结果通过胎儿心脏 MRI 进行评估。

方法

开发了一种 MRI 条件 DUS 设备,以从胎儿心脏获得门控信号。在 1.5T 和 3T 下使用 B 场图和梯度回波图像评估 MRI 兼容性。在 MRI 室外,对 15 例胎儿进行了 DUS 设备检测胎儿心脏运动进行心脏门控的质量和灵敏度评估。采用动态胎儿心脏体模评估 1.5T 和 3T 时电磁干扰对 DUS 设备和门控信号的失真。在首次体内经验中,在 2 例胎儿中在 1.5T 和 3T 下采集四腔心视图的动态胎儿心脏图像。

结果

在 1.5T 和 3T 下,有和没有 DUS 设备时 B 场和信号强度的最大变化均<6.5%。DUS 设备检测胎儿心跳的灵敏度为 99.1%。使用胎儿心脏体模验证 DUS 设备时,在 1.5T 或 3T 下未发现电磁干扰,与模拟心脏频率具有高度相关性。成功应用胎儿心脏电影图像,显示出良好的图像质量。

结论

开发了一种 MR 条件 DUS 门控设备,并对其在不同场强下的安全性、兼容性和可靠性进行了评估。在初步经验中,DUS 设备成功应用于 1.5T 和 3T 下的体内胎儿心脏成像。