使用带有基准标记的肝脏立体定向体放射治疗的不同机架旋转速度获得的四维锥形束 CT 的图像质量。

Image quality of four-dimensional cone-beam computed tomography obtained at various gantry rotation speeds for liver stereotactic body radiation therapy with fiducial markers.

机构信息

Department of Radiological Technology, Kumamoto University Hospital, Kumamoto, Japan; Graduate School of Health Sciences, Kumamoto University, Kumamoto, Japan.

Department of Health Sciences, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan.

出版信息

Phys Med. 2018 Jan;45:19-24. doi: 10.1016/j.ejmp.2017.11.020. Epub 2017 Dec 19.

Abstract

In this study, qualities of 4D cone-beam CT (CBCT) images obtained using various gantry rotation speeds (GRSs) for liver stereotactic body radiation therapy (SBRT) with fiducial markers were quantitatively evaluated. Abdominal phantom containing a fiducial marker was moved along a sinusoidal waveform, and 4D-CBCT images were acquired with GRSs of 50-200° min. We obtained the 4D-CBCT projection data from six patients who underwent liver SBRT and generated 4D-CBCT images at GRSs of 67-200° min, by varying the number of projection data points. The image quality was evaluated based on the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and structural similarity index (SSIM). The fiducial marker positions with different GRSs were compared with the setup values and a reference position in the phantom and clinical studies, respectively. The root mean square errors (RMSEs) were calculated relative to the reference positions. In the phantom study, the mean SNR, CNR, and SSIM decreased from 37.6 to 10.1, from 39.8 to 10.1, and from 0.9 to 0.7, respectively, as the GRS increased from 50 to 200° min. The fiducial marker positions were within 2.0 mm at all GRSs. Similarly, in the clinical study, the mean SNR, CNR, and SSIM decreased from 50.4 to 13.7, from 24.2 to 6.0, and from 0.92 to 0.73, respectively. The mean RMSEs were 2.0, 2.1, and 3.6 mm for the GRSs of 67, 100, and 200° min, respectively. We conclude that GRSs of 67 and 85° min yield images of acceptable quality for 4D-CBCT in liver SBRT with fiducial markers.

摘要

在这项研究中,定量评估了使用不同机架旋转速度(GRS)获得的带有基准标记的肝脏立体定向体放射治疗(SBRT)的 4D 锥形束 CT(CBCT)图像的质量。含有基准标记的腹部体模沿着正弦波形移动,以 50-200°/min 的 GRS 采集 4D-CBCT 图像。我们从接受肝脏 SBRT 的六名患者中获得了 4D-CBCT 投影数据,并通过改变投影数据点数,以 67-200°/min 的 GRS 生成了 4D-CBCT 图像。根据信噪比(SNR)、对比噪声比(CNR)和结构相似性指数(SSIM)评估图像质量。使用不同 GRS 的基准标记位置与体模和临床研究中的设置值和参考位置进行了比较。相对于参考位置计算了均方根误差(RMSE)。在体模研究中,随着 GRS 从 50 增加到 200°/min,平均 SNR、CNR 和 SSIM 分别从 37.6 降低到 10.1、从 39.8 降低到 10.1 和从 0.9 降低到 0.7。在所有 GRS 下,基准标记位置均在 2.0mm 以内。同样,在临床研究中,平均 SNR、CNR 和 SSIM 分别从 50.4 降低到 13.7、从 24.2 降低到 6.0 和从 0.92 降低到 0.73。GRS 分别为 67、100 和 200°/min 时,平均 RMSE 分别为 2.0、2.1 和 3.6mm。我们得出结论,67 和 85°/min 的 GRS 可产生具有可接受质量的带有基准标记的肝脏 SBRT 的 4D-CBCT 图像。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验