Abstract

As a cost-effective, clinician-friendly gait assessment tool, the Kinect v2 sensor may be effective for assessing lower extremity joint kinematics. This study aims to examine the validity of time series kinematical data as measured by the Kinect v2 on a flatland for gait assessment. In this study, 51 healthy subjects walked on a flatland while kinematic data were extracted concurrently using the Kinect and Vicon systems. The kinematic outcomes comprised the hip and knee joint angles. Parallel translation of Kinect data obtained throughout the gait cycle was performed to minimize the differences between the Kinect and Vicon data. The ensemble curves of the hip and knee joint angles were compared to investigate whether the Kinect sensor can consistently and accurately assess lower extremity joint motion throughout the gait cycle. Relative consistency was assessed using Pearson correlation coefficients. Joint angles measured by the Kinect v2 followed the trend of the trajectories made by the Vicon data in both the hip and knee joints in the sagittal plane. The trajectories of the hip and knee joint angles in the frontal plane differed between the Kinect and Vicon data. We observed moderate to high correlation coefficients of 20%-60% of the gait cycle, and the largest difference between Kinect and Vicon data was 4.2°. Kinect v2 time series kinematical data obtained on the flatland are validated if the appropriate correction procedures are performed. Future studies are warranted to examine the reproducibility and systematic bias of the Kinect v2.

摘要

作为一种经济高效、对临床医生友好的步态评估工具,Kinect v2传感器可能对评估下肢关节运动学有效。本研究旨在检验通过Kinect v2在平地上测量的时间序列运动学数据用于步态评估的有效性。在本研究中,51名健康受试者在平地上行走,同时使用Kinect和Vicon系统同步提取运动学数据。运动学结果包括髋关节和膝关节角度。对整个步态周期获得的Kinect数据进行平行平移,以尽量减少Kinect和Vicon数据之间的差异。比较髋关节和膝关节角度的整体曲线,以研究Kinect传感器能否在整个步态周期中持续、准确地评估下肢关节运动。使用Pearson相关系数评估相对一致性。在矢状面中,Kinect v2测量的髋关节和膝关节角度均遵循Vicon数据所形成轨迹的趋势。在额状面中,Kinect和Vicon数据的髋关节和膝关节角度轨迹有所不同。我们观察到在步态周期的20%-60%范围内,相关系数为中度到高度,Kinect和Vicon数据之间的最大差异为4.2°。如果执行适当的校正程序,在平地上获得的Kinect v2时间序列运动学数据是有效的。未来的研究有必要检验Kinect v2的可重复性和系统偏差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验