Nawale Jaywant M, Patil Sandip N, Nalawade Digvijay D, Borikar Nikhil A, Sonawane Bhushan S, Chaurasia Ajay S
Department of Cardiology, TNMC & BYL NAIR Ch. Hospital Mumbai, India.
Department of Cardiology, TNMC & BYL NAIR Ch. Hospital Mumbai, India.
Indian Pacing Electrophysiol J. 2018 Jul-Aug;18(4):148-149. doi: 10.1016/j.ipej.2018.02.004. Epub 2018 Mar 2.
Implantation of temporary pacemaker lead is commonly performed procedure and is usually safe, but can sometimes develop rare and serious complication like intracardiac lead knotting which may require challenging retrieval techniques. We report a case of successful percutaneous retrieval of unusually knotted right internal jugular venous temporary pacing lead via left femoral transvenous approach using snare over a long sheath after cutting the electrode proximally and thus avoiding any surgical intervention.
临时起搏器导线植入是一种常见的操作,通常是安全的,但有时可能会出现罕见且严重的并发症,如心内导线打结,这可能需要采用具有挑战性的取出技术。我们报告了一例通过左股静脉途径,在近端切断电极后,使用圈套器套在长鞘管上,成功经皮取出异常打结的右颈内静脉临时起搏导线的病例,从而避免了任何外科干预。