麻醉犬中加速度肌动图与机械肌动图之间的偏差、一致性界限和百分比误差。

Bias, limits of agreement, and percent errors between acceleromyography and mechanomyography in anesthetized dogs.

作者信息

Sakai D M, Romano M, Tseng C T, Flanders A, Martin-Flores M

机构信息

Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853, USA.

Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853, USA.

出版信息

Vet J. 2018 Mar;233:3-7. doi: 10.1016/j.tvjl.2017.12.020. Epub 2017 Dec 29.

Abstract

This study compared measurements of neuromuscular function with mechanomyography (MMG) and acceleromyography (AMG) in nine anesthetized dogs receiving 0.1mg/kg vecuronium intravenously. Train-of-four (TOF) stimulation was applied to each ulnar nerve every 15s. The resulting amplitude of the first twitch (T1) and the TOF ratio were measured with both monitors. The baseline TOF ratio (prior to vecuronium), onset time (time of injection to T1<5%), recovery index (time between T1 values of 25% and 75%) and duration of neuromuscular block (injection to TOF 0.9) were recorded. The MMG TOF ratios when the AMG first reached 0.7 (AMG 0.7) and 0.9 (AMG 0.9) during recovery were also recorded. Values were compared with paired tests and individual errors>25% between monitors were identified for each dog. Bias, limits of agreement (LOA) and percentage error (PE) between methods were calculated from Bland-Altman plots for T1 and TOF ratio for the complete data set, and for TOF≥0.7 during recovery. There were no statistical differences in baseline TOF ratio, onset, recovery index, duration, AMG 0.7 and AMG 0.9. Individual errors>25% were evident in onset, recovery index, AMG 0.7 and AMG 0.9. Overall, T1 and TOF ratio had a small bias, wide LOA and PE>100%. Percent error was reduced to 30% when TOF≥0.7 was analyzed. Although there were no statistical differences between MMG and AMG in any variable of interest, individual discrepancies, wide LOA and high PE suggest that these monitors should not be used interchangeably for serial measurements on the same animals.

摘要

本研究比较了在9只静脉注射0.1mg/kg维库溴铵的麻醉犬中,使用肌机械图(MMG)和加速度肌电图(AMG)测量神经肌肉功能的情况。每隔15秒对每条尺神经施加四个成串刺激(TOF)。用两种监测仪测量由此产生的第一个肌颤搐(T1)的幅度和TOF比值。记录基线TOF比值(维库溴铵给药前)、起效时间(注射至T1<5%的时间)、恢复指数(T1值在25%和75%之间的时间)以及神经肌肉阻滞持续时间(注射至TOF 0.9)。还记录了恢复过程中AMG首次达到0.7(AMG 0.7)和0.9(AMG 0.9)时的MMG TOF比值。采用配对检验比较数值,并确定每只犬监测仪之间个体误差>25%的情况。根据Bland-Altman图,针对完整数据集以及恢复过程中TOF≥0.7时的T1和TOF比值,计算两种方法之间的偏差、一致性界限(LOA)和百分比误差(PE)。基线TOF比值、起效时间、恢复指数、持续时间、AMG 0.7和AMG 0.9之间无统计学差异。在起效时间、恢复指数、AMG 0.7和AMG 0.9方面,个体误差>25%较为明显。总体而言,T1和TOF比值偏差较小、LOA较宽且PE>100%。分析TOF≥0.7时,百分比误差降至30%。尽管在任何感兴趣的变量上MMG和AMG之间无统计学差异,但个体差异、较宽的LOA和较高的PE表明,这些监测仪不应在同一动物上进行连续测量时互换使用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验