KIOS Research Centre, Department of Electrical and Computer Engineering, University of Cyprus, Nicosia, Cyprus.
Physiol Meas. 2018 Apr 26;39(4):044004. doi: 10.1088/1361-6579/aab513.
Critically ill neonates and infants might particularly benefit from continuous chest electrical impedance tomography (EIT) monitoring at the bedside. In this study a textile 32-electrode interface for neonatal EIT examination has been developed and tested to validate its clinical performance. The objectives were to assess ease of use in a clinical setting, stability of contact impedance at the electrode-skin interface and possible adverse effects.
Thirty preterm infants (gestational age: 30.3 ± 3.9 week (mean ± SD), postnatal age: 13.8 ± 28.2 d, body weight at inclusion: 1727 ± 869 g) were included in this multicentre study. The electrode-skin contact impedances were measured continuously for up to 3 d and analysed during the initial 20-min phase after fastening the belt and during a 10 h measurement interval without any clinical interventions. The skin condition was assessed by attending clinicians.
Our findings imply that the textile electrode interface is suitable for long-term neonatal chest EIT imaging. It does not cause any distress for the preterm infants or discomfort. Stable contact impedance of about 300 Ohm was observed immediately after fastening the electrode belt and during the subsequent 20 min period. A slight increase in contact impedance was observed over time. Tidal variation of contact impedance was less than 5 Ohm.
The availability of a textile 32-electrode belt for neonatal EIT imaging with simple, fast, accurate and reproducible placement on the chest strengthens the potential of EIT to be used for regional lung monitoring in critically ill neonates and infants.
危重新生儿和婴儿可能特别受益于床边连续胸部电阻抗断层成像(EIT)监测。本研究开发并测试了一种用于新生儿 EIT 检查的纺织 32 电极接口,以验证其临床性能。目的是评估在临床环境中的易用性、电极-皮肤界面接触阻抗的稳定性以及可能的不良反应。
这项多中心研究纳入了 30 名早产儿(胎龄:30.3±3.9 周(均值±标准差),生后年龄:13.8±28.2 天,纳入时体重:1727±869g)。连续测量电极-皮肤接触阻抗长达 3 天,并在固定腰带后的最初 20 分钟阶段和 10 小时测量间隔内(无任何临床干预)进行分析。临床医生评估皮肤状况。
我们的研究结果表明,纺织电极接口适合长期新生儿胸部 EIT 成像。它不会给早产儿带来任何不适或不适。在固定电极带后,立即观察到约 300 欧姆的稳定接触阻抗,并在随后的 20 分钟期间保持稳定。随着时间的推移,接触阻抗略有增加。接触阻抗的潮气量变化小于 5 欧姆。
具有简单、快速、准确和可重复放置在胸部的 32 电极纺织带的可用性,增强了 EIT 用于监测危重新生儿和婴儿肺部区域的潜力。