Industrial Technology Research Institute, Hsinchu 310, Taiwan.
Department of Electrical Engineering, National Cheng Kung University, Tainan 701, Taiwan.
Sensors (Basel). 2021 Oct 13;21(20):6789. doi: 10.3390/s21206789.
Electrical impedance tomography (EIT), a noninvasive and radiation-free medical imaging technique, has been used for continuous real-time regional lung aeration. However, adhesive electrodes could cause discomfort and increase the risk of skin injury during prolonged measurement. Additionally, the conductive gel between the electrodes and skin could evaporate in long-term usage and deteriorate the signal quality. To address these issues, in this work, textile electrodes integrated with a clothing belt are proposed to achieve EIT lung imaging along with a custom portable EIT system. The simulation and experimental results have verified the validity of the proposed portable EIT system. Furthermore, the imaging results of using the proposed textile electrodes were compared with commercial electrocardiogram electrodes to evaluate their performance.
电阻抗断层成像(EIT)是一种非侵入性且无辐射的医学成像技术,已被用于连续实时区域性肺部充气成像。然而,在长时间测量过程中,粘性电极可能会引起不适并增加皮肤损伤的风险。此外,电极和皮肤之间的导电凝胶在长期使用过程中可能会蒸发,从而降低信号质量。为了解决这些问题,本工作提出了一种与服装带集成的纺织电极,以实现 EIT 肺部成像以及定制的便携式 EIT 系统。模拟和实验结果验证了所提出的便携式 EIT 系统的有效性。此外,还使用所提出的纺织电极的成像结果与商用心电图电极进行了比较,以评估它们的性能。