Stein P D, Sabbah H N, Albert D E, Snyder J E
Med Instrum. 1987 Jun;21(3):177-82.
A new continuous-wave Doppler device is described, which has the capability of measuring peak aortic blood velocity and acceleration noninvasively in the ascending aorta of patients. To test the accuracy of the device, blood velocity and acceleration in the ascending aorta were compared with measurements obtained using an electromagnetic flowmeter in 16 open-chest anesthetized dogs. The Doppler probe was hand held directly on the aorta. Aortic flow was measured with a cuff electromagnetic flow transducer placed at the root of the aorta. Isoproterenol and propranolol, sometimes in combination with lidocaine, were administered intravenously to augment or reduce left ventricular contractile performance. Values of peak velocity, measured with the Doppler, corresponded closely to values measured with the electromagnetic flowmeter (r = 0.95). Values of peak acceleration also corresponded closely with the electromagnetic flow measurements (r = 0.96). The results indicate that valid measurements of blood acceleration in the ascending aorta, as well as blood velocity, can be obtained with continuous-wave Doppler.
本文描述了一种新型连续波多普勒装置,它能够在患者升主动脉中无创测量主动脉血流峰值速度和加速度。为测试该装置的准确性,将16只开胸麻醉犬升主动脉中的血流速度和加速度与使用电磁流量计获得的测量值进行了比较。多普勒探头直接手持置于主动脉上。用置于主动脉根部的袖带式电磁流量传感器测量主动脉血流。静脉注射异丙肾上腺素和普萘洛尔,有时联合利多卡因,以增强或降低左心室收缩功能。用多普勒测得的峰值速度值与用电磁流量计测得的值密切相关(r = 0.95)。峰值加速度值也与电磁流量测量值密切相关(r = 0.96)。结果表明,使用连续波多普勒能够获得升主动脉中血流加速度以及血流速度的有效测量值。