Ren Liqiang, McCollough Cynthia H, Yu Lifeng
Department of Radiology, Mayo Clinic, Rochester, MN.
Proc SPIE Int Soc Opt Eng. 2018 Mar;10573. doi: 10.1117/12.2294953.
In order to perform material decomposition for a three-material mixture, dual-energy CT (DECT) has to incorporate an additional condition, typically the prior information related to certain physical constraints such as volume or mass conservation. With the introduction of photon-counting CT and other multi-energy CT (MECT) platform, more than 2 energy bins can be simultaneously acquired, which in principle can solve a three-material problem without the need of additional prior information. The purpose of this work was to investigate the impact of prior information on noise and bias properties of three-material decomposition in both DECT and MECT, and to evaluate if the prior information is still needed in MECT. Computer simulation studies were performed to compare basis image noise and quantification accuracy among DECT with prior information, and MECT with/without prior information. For given spectral configurations, the simulation results showed that significant noise reductions can be achieved in all the basis material images when prior information was included in the material decomposition process. Compared to DECT with prior information, MECT (=3) with prior information had slightly better noise performance due to additional beam measurement and well preserved spectral separation. In addition, when wrong prior information ([-2.0%, 2.0%]) was intentionally introduced, the quantification accuracy evaluated by root-mean-square-error (RMSR) using MECT with prior information was less than 1.5mg/cc for gadolinium quantification and 1.2mg/cc for iodine quantification.
为了对三物质混合物进行物质分解,双能CT(DECT)必须纳入一个附加条件,通常是与某些物理约束(如体积或质量守恒)相关的先验信息。随着光子计数CT和其他多能CT(MECT)平台的引入,可以同时获取超过2个能量区间的数据,原则上这可以在无需额外先验信息的情况下解决三物质问题。本研究的目的是探讨先验信息对DECT和MECT中三物质分解的噪声和偏差特性的影响,并评估MECT中是否仍需要先验信息。进行了计算机模拟研究,以比较有先验信息的DECT以及有/无先验信息的MECT之间的基图像噪声和定量准确性。对于给定的光谱配置,模拟结果表明,当在物质分解过程中纳入先验信息时,所有基物质图像中的噪声都能显著降低。与有先验信息的DECT相比,有先验信息的MECT(=3)由于额外的束流测量和良好保存的光谱分离,具有略好的噪声性能。此外,当故意引入错误的先验信息([-2.0%,2.0%])时,使用有先验信息的MECT通过均方根误差(RMSR)评估的定量准确性对于钆定量小于1.5mg/cc,对于碘定量小于1.2mg/cc。