Biologic fixation of ligament prostheses and augmentations. An evaluation of bone ingrowth in the dog.

作者信息

Arnoczky S P, Torzilli P A, Warren R F, Allen A A

机构信息

Laboratory of Comparative Orthopaedics, Hospital for Special Surgery, New York, NY 10021.

出版信息

Am J Sports Med. 1988 Mar-Apr;16(2):106-12. doi: 10.1177/036354658801600204.

Abstract

The biologic fixation (bone ingrowth) of three prosthetic ligament devices (bovine xenograft Xenotech Laboratories, Inc., Irvine, CA; Gore-Tex, W. L. Gore and Associates, Flagstaff, AZ; and knitted dacron, Stryker, Kalamazoo, MI) and one ligament augmentation device (LAD, braided polypropylene, 3M, St. Paul, MN) were evaluated in vivo. The devices were placed unstressed, in an extraarticular location in the cortical-cancellous bone of the proximal femur and humerus of dogs. Six months following implantation tissue ingrowth was evaluated using high resolution radiography, routine histology, scanning electron microscopy, and biomechanical tensile tests to measure the ultimate (maximum) pull-out strength of the devices. The knitted dacron and Gore-Tex prosthesis showed marked ingrowth of trabecular bone into the interstices of the device with evidence of bone growth around individual fibers of the prosthesis. The bovine xenograft demonstrated bone immediately adjacent to the material with evidence of spicular ingrowth of bone into the natural clefts of the prosthesis. The LAD showed no evidence of bone ingrowth and demonstrated a fibrous tissue interface between the prosthesis and surrounding trabecular bone. Six months following implantation the pull-out strengths of the devices were as follows: knitted dacron, 445.7 +/- 151.0N; Gore-Tex, 438.1 +/- 131.94N; bovine xenograft, 332.8 +/- 64.2N; and LAD, 78.4 +/- 47.9N. Statistical analysis revealed no significant difference between the pull-out strengths of the knitted dacron, Gore-Tex, and bovine xenograft.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

在体内评估了三种人工韧带装置(牛异种移植物,Xenotech Laboratories公司,加利福尼亚州欧文市;戈尔特斯,W.L.戈尔联合公司,亚利桑那州弗拉格斯塔夫市;针织涤纶,史赛克公司,密歇根州卡拉马祖市)和一种韧带增强装置(LAD,编织聚丙烯,3M公司,明尼苏达州圣保罗市)的生物固定(骨长入)情况。将这些装置无张力地放置在犬近端股骨和肱骨皮质-松质骨的关节外位置。植入6个月后,使用高分辨率射线照相术、常规组织学、扫描电子显微镜和生物力学拉伸试验评估组织长入情况,以测量装置的极限(最大)拔出强度。针织涤纶和戈尔特斯假体显示小梁骨显著长入装置间隙,假体个别纤维周围有骨生长迹象。牛异种移植物显示材料紧邻骨,有骨针状长入假体天然裂隙的迹象。LAD没有骨长入迹象,假体与周围小梁骨之间显示为纤维组织界面。植入6个月后,装置的拔出强度如下:针织涤纶,445.7±151.0N;戈尔特斯,438.1±131.94N;牛异种移植物,332.8±64.2N;LAD,78.4±47.9N。统计分析显示针织涤纶、戈尔特斯和牛异种移植物的拔出强度之间无显著差异。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验