Suppr超能文献

定量计算机断层扫描(QCT)测量椎体骨矿物质密度,有无校准体模:两种不同软件解决方案的比较。

Vertebral Bone Mineral Density Measured by Quantitative Computed Tomography With and Without a Calibration Phantom: A Comparison Between 2 Different Software Solutions.

机构信息

Department of Internal Medicine, Hospital Unit West, Herning, Denmark.

Department of Clinical Engineering, Aarhus University Hospital, Aarhus, Denmark.

出版信息

J Clin Densitom. 2018 Jul-Sep;21(3):367-374. doi: 10.1016/j.jocd.2017.12.003. Epub 2018 Mar 27.

Abstract

Quantitative computed tomography (CT) can be used to quantify bone mineral density (BMD) in the spine from clinical CT scans. We aimed to determine agreement and precision of BMD measurements by 2 different methods: phantom-less internal tissue calibration and asynchronous phantom-based calibration in a cohort of patients with chronic kidney disease (CKD). Patients with CKD were recruited for CT angiography of the chest, abdomen, and pelvis. BMD was analyzed by 2 different software solutions using different calibration techniques; phantom-based by QCT Pro (Mindways Inc.) and phantom-less by Extended Brilliance Workspace (Philips Healthcare). Intraoperator reanalysis was performed on 53 patients (36%) for both methods. An interoperator reanalysis on 30 patients (20%) using the phantom-based method and 29 patients (19%) using the phantom-less method was made. XY- and Bland-Altman plots were used to evaluate method agreement. Phantom-based measured BMD was systematically higher than phantom-less measured BMD. Despite a small absolute difference of 3.3 mg/cm (CI: -0.2-6.9 mg/cm) and a relative difference of 5.1% (CI: 2.2%-8.1%), interindividual differences were large, as seen by a wide prediction interval (PI: -47-40 mg/cm). The Bland-Altman plot showed no systematic bias, apart from 5 outliers. Intraoperator variability was high for the phantom-less method (5.8%) compared to the phantom-based (0.8%) and the interoperator variability was also high for the phantom-less method (5.8%) compared to the phantom-based (1.8%). Despite high correlation between methods, the between-method difference on an individual level showed great variability. Our results suggest agreement between these 2 methods is insufficient to allow them to be used interchangeably in patients with CKD.

摘要

定量计算机断层扫描(CT)可用于从临床 CT 扫描中定量测量脊柱的骨矿物质密度(BMD)。我们旨在通过两种不同的方法:无模体内部组织校准和慢性肾脏病(CKD)患者的异步模体校准,确定两种方法测量 BMD 的一致性和精度。招募 CKD 患者进行胸部、腹部和骨盆 CT 血管造影。使用两种不同的校准技术,两种不同的软件解决方案分析 BMD;QCT Pro(Mindways Inc.)使用基于模体的方法,Extended Brilliance Workspace(Philips Healthcare)使用无模体方法。两种方法都对 53 名患者(36%)进行了操作者内再分析。使用基于模体的方法对 30 名患者(20%)和使用无模体的方法对 29 名患者(19%)进行了操作者间再分析。使用 XY-和 Bland-Altman 图评估方法的一致性。基于模体的测量 BMD 系统地高于无模体的测量 BMD。尽管绝对差异较小(3.3mg/cm(CI:-0.2-6.9mg/cm)和相对差异较小(5.1%(CI:2.2%-8.1%)),但个体间差异较大,预测区间较宽(PI:-47-40mg/cm)。Bland-Altman 图显示除 5 个离群值外,没有系统偏差。无模体方法的操作者内变异性较高(5.8%),而基于模体的方法较低(0.8%),无模体方法的操作者间变异性也较高(5.8%),而基于模体的方法较低(1.8%)。尽管两种方法之间相关性较高,但个体水平上的方法间差异显示出很大的可变性。我们的结果表明,这两种方法之间的一致性不足以允许它们在 CKD 患者中互换使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验