Osteoporosis is a systemic bone disease which can increase the risk of osteoporotic fractures. Dual-energy X-ray absorptiometry (DXA) is considered as the clinical standard for diagnosing osteoporosis by detecting the bone mineral density (BMD) in patients, but it has flaws in distinguishing between calcification and other degenerative diseases, thus leading to inaccurate BMD levels in subjects. Mindways quantitative computed tomography (Mindways QCT) is a classical QCT system. Similar to DXA, Mindways QCT can directly present the density of trabecular bone, vascular or tissue calcification; therefore, it is more accurate and sensitive than DXA and has been widely applied in clinic to evaluate osteoporosis. iCare QCT osteodensitometry was a new phantom-based QCT system, recently developed by iCare Inc. (China). It has been gradually applied in clinic by its superiority of taking 3-dimensional BMD of bone and converting BMD values to T value automatically. This study aimed at evaluating the osteoporosis detection rate of iCare QCT, compared with synchronous Mindways QCT (USA).
In this study, 131 patients who underwent hip phantom-based CT scan were included. Bone mineral density (BMD) of the unified region of interests (ROI) defined at the European spine phantom (ESP, German QRM) including L1 (low), L2 (medium), and L3 (high) vertebral bodies was detected for QCT quality control and horizontal calibration. Every ESP scan were taken for 10 times, and the mean BMD values measured by iCare QCT and Mindways QCT were compared. Hip CT scan was conducted with ESP as calibration individually. T-scores gained from iCare QCT and Mindways QCT were analyzed with Pearson correlation test. The detection rates of osteoporosis were compared between iCare QCT and Mindways QCT. The unified region of interests (ROI) was delineated in the QCT software.
The results showed that there was no significant difference between iCare QCT and Mindways QCT in the evaluation of L1, L2, and L3 vertebrae bodies in ESP. A strong correlation between iCare QCT and Mindways QCT in the assessment of hip T-score was found. It was illustrated that iCare QCT had a higher detection rate of osteoporosis with the assessment of hip T-score than Mindways QCT did. In patients < 50 years subgroup, the detection rate of osteoporosis with iCare QCT and Mindways QCT was equal. In patients ≥ 50 years subgroup, the detection rate of osteoporosis with iCare QCT (35/92, 38.0%) was higher than that with Mindways QCT. In female subgroup, the detection rate of osteoporosis with iCare QCT was significantly higher than Mindways QCT. In male subgroup, the detection rate of osteoporosis with iCare QCT was also markedly higher than Mindways QCT. The detection rate of osteoporosis by iCare QCT was higher than Mindways QCT with hip bone assessment. Of course, the results of the present study remain to be further verified by multicenter studies in the future.
骨质疏松症是一种全身性骨骼疾病,会增加骨质疏松性骨折的风险。双能 X 射线吸收法(DXA)通过检测患者的骨矿物质密度(BMD),被认为是诊断骨质疏松症的临床标准,但它在区分钙化和其他退行性疾病方面存在缺陷,因此导致 BMD 水平在受检者中不准确。Mindways 定量计算机断层扫描(Mindways QCT)是一种经典的 QCT 系统。与 DXA 类似,Mindways QCT 可以直接呈现骨小梁、血管或组织钙化的密度;因此,它比 DXA 更准确和敏感,并已广泛应用于临床评估骨质疏松症。iCare QCT 骨密度仪是 iCare Inc.(中国)最近开发的一种新的基于体模的 QCT 系统。它具有获取骨骼 3D BMD 并自动将 BMD 值转换为 T 值的优势,因此在临床应用中逐渐得到应用。本研究旨在评估 iCare QCT 检测骨质疏松症的能力,与同步的 Mindways QCT(美国)进行比较。
本研究纳入了 131 例接受髋关节体模 CT 扫描的患者。使用欧洲脊柱体模(ESP,德国 QRM)定义的统一感兴趣区(ROI)检测骨矿物质密度(BMD),包括 L1(低)、L2(中)和 L3(高)椎体,用于 QCT 质量控制和水平校准。每个 ESP 扫描进行 10 次,比较 iCare QCT 和 Mindways QCT 测量的平均 BMD 值。单独使用 ESP 进行髋关节 CT 扫描校准。使用 Pearson 相关检验分析 iCare QCT 和 Mindways QCT 获得的 T 评分。比较 iCare QCT 和 Mindways QCT 对骨质疏松症的检出率。在 QCT 软件中划定统一的感兴趣区(ROI)。
结果显示,iCare QCT 和 Mindways QCT 在 ESP 中评估 L1、L2 和 L3 椎体时,结果无显著差异。iCare QCT 和 Mindways QCT 在髋关节 T 评分评估方面具有很强的相关性。表明 iCare QCT 对髋关节 T 评分的评估具有更高的骨质疏松症检出率。在<50 岁的患者亚组中,iCare QCT 和 Mindways QCT 对骨质疏松症的检出率相等。在≥50 岁的患者亚组中,iCare QCT(35/92,38.0%)对骨质疏松症的检出率高于 Mindways QCT。在女性亚组中,iCare QCT 对骨质疏松症的检出率明显高于 Mindways QCT。在男性亚组中,iCare QCT 对骨质疏松症的检出率也明显高于 Mindways QCT。iCare QCT 对髋关节骨评估的骨质疏松症检出率高于 Mindways QCT。当然,本研究的结果还需要进一步通过未来的多中心研究来验证。