Garza-Leon Manuel, Plancarte-Lozano Eduardo, Valle-Penella Agustín Del, Guzmán-Martínez María de Lourdes, Villarreal-González Andrés
Fundación Destellos de Luz, Universidad de Monterrey. Monterrey, N.L. México.
Dirección de Ciencias Médicas de la División de Ciencias de la Salud, Universidad de Monterrey. Monterrey, N.L. México.
Cir Cir. 2018;86(1):50-55. doi: 10.24875/CIRU.M18000002.
Corneal pachymetry is widely used for refractive surgery and follow up in keratoconus, accurate measurement is essential for a safe surgery.
To assess intraobserver reliability of central corneal thickness (CCT) measurements using optical low-coherence reflectometry (OLCR) technology and its agreement with ultrasonic pachymeter (US).
Randomized and prospective comparative evaluation of diagnostic technology. One randomly healthy eye of subjects was scanned three times with both devices. Intraobserver within-subject standard deviation (Sw), coefficient of variation (CVw) and intraclass correlation coefficient (ICC) were obtained for reliability analysis; for study agreement, data were analyzed using the paired-sample t test and the Bland-Altman LoA method. The mean of three scans of each equipment was used to assess the LoA.
The study enrolled 30 eyes of 30 subjects with average age of 28.70 ± 8.06 years. For repeatability, the Sw were 3.41 and 5.96 µ, the intraobserver CVw was 2 and 4% and ICC 0.991 and 0.988, for OLCR and US respectively. The mean CCT difference between OLCR and US was 8.90 ± 9.03 µ (95% confidence interval: 5.52-2.27 µ), and the LoA was 35.40 µ.
OLCR technology provided reliable intraobserver CCT measurements. Both pachymetry measurements may be used interchangeably with minimum calibration adjustment.
角膜测厚在屈光手术及圆锥角膜随访中广泛应用,准确测量对安全手术至关重要。
评估使用光学低相干反射ometry(OLCR)技术测量中央角膜厚度(CCT)的观察者内可靠性及其与超声测厚仪(US)的一致性。
对诊断技术进行随机前瞻性比较评估。用两种设备对受试者的一只随机选取的健康眼进行三次扫描。获取观察者内受试者标准差(Sw)、变异系数(CVw)和组内相关系数(ICC)进行可靠性分析;对于研究一致性,使用配对样本t检验和Bland-Altman LoA方法分析数据。用每种设备三次扫描的平均值评估LoA。
该研究纳入了30名受试者的30只眼,平均年龄为28.70±8.06岁。对于重复性,OLCR和US的Sw分别为3.41和5.96µ,观察者内CVw分别为2%和4%,ICC分别为0.991和0.988。OLCR和US之间的平均CCT差值为8.90±9.03µ(95%置信区间:5.52 - 2.27µ),LoA为35.40µ。
OLCR技术提供了可靠的观察者内CCT测量值。两种测厚测量在进行最小校准调整后可互换使用。