Suppr超能文献

标准超声角膜测厚法、角膜地形图、眼压-角膜测厚法及眼前节光学相干断层扫描测量中央角膜厚度的比较

Comparison of Central Corneal Thickness Measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomography.

作者信息

González-Pérez Javier, Queiruga Piñeiro Juan, Sánchez García Ángelx, González Méijome José Manuel

机构信息

a Ocular Surface and Contact Lens Research Laboratory, Faculty of Optometry , University of Santiago de Compostela , Santiago de Compostela , Spain.

b Instituto Oftalmológico Fernández-Vega , Oviedo , Spain.

出版信息

Curr Eye Res. 2018 Jul;43(7):866-872. doi: 10.1080/02713683.2018.1461910. Epub 2018 Apr 13.

Abstract

PURPOSE

To compare central corneal thickness (CCT) measured by standard ultrasound pachymetry (USP), and three non-contact devices in healthy eyes.

METHODS

A cross-sectional study of CCT measurement in 52 eyes of 52 healthy volunteers was done by a single examiner at Ocular Surface and Contact Lens Laboratory. Three consecutive measurements were done by standard USP, non-contact tono-pachymeter, Pentacam corneal topographer, and Anterior Segment Optical Coherence Tomography (AS-OCT). The mean values were used for assessment. The results were compared using multivariate ANOVA, linear regression, and Pearson correlation. Agreement among the devices was analyzed using mean differences and Bland-Altman analysis with 95% limits of agreement (LoA). Finally, reliability was analyzed using intraclass correlation coefficient (ICC).

RESULTS

Mean CCT by ultrasound pachymeter, tono-pachymeter, corneal topographer and AS-OCT were 558.9 ± 31.2 µm, 525.8 ± 43.1 µm, 550.4 ± 30.5 µm, and 545.9 ± 30.5 µm respectively. There was a significant positive correlation between AS-OCT and USP (Pearson correlation = 0.957, p < 0.001), corneal topography and USP (Pearson correlation = 0.965, p < 0.001), and corneal topography and AS-OCT (Pearson correlation = 0.965, p < 0.001). There was a lower correlation between CT-1P tono-pachymeter and the other three modalities. Intraclass correlation coefficients show an excellent reliability between pairs except for CT-1P against the other three instruments that were found moderate.

CONCLUSIONS

CT-1P tono-pachymeter underestimates CCT measurements compared to Scheimpflug system, AS-OCT device, and USP. Mean CCT among USP, Pentacam and AS-OCT were comparable and had significant linear correlations. In clinical practice, these three modalities could be interchangeable in healthy patients.

摘要

目的

比较标准超声角膜测厚法(USP)与三种非接触式设备在健康眼中测量的中央角膜厚度(CCT)。

方法

在眼表与隐形眼镜实验室,由一名检查者对52名健康志愿者的52只眼睛进行CCT测量的横断面研究。通过标准USP、非接触眼压角膜厚度仪、Pentacam角膜地形图仪和眼前节光学相干断层扫描(AS-OCT)进行连续三次测量。使用平均值进行评估。采用多变量方差分析、线性回归和Pearson相关性分析比较结果。使用平均差异和Bland-Altman分析以及95%一致性界限(LoA)分析设备之间的一致性。最后,使用组内相关系数(ICC)分析可靠性。

结果

超声角膜测厚仪、眼压角膜厚度仪、角膜地形图仪和AS-OCT测量的平均CCT分别为558.9±31.2μm、525.8±43.1μm、550.4±30.5μm和545.9±30.5μm。AS-OCT与USP之间存在显著正相关(Pearson相关性=0.957,p<0.001),角膜地形图与USP之间存在显著正相关(Pearson相关性=0.965,p<0.001),角膜地形图与AS-OCT之间存在显著正相关(Pearson相关性=0.965,p<0.001)。CT-1P眼压角膜厚度仪与其他三种测量方式之间的相关性较低。组内相关系数显示,除CT-1P与其他三种仪器之间为中等可靠性外,各对设备之间具有极好的可靠性。

结论

与Scheimpflug系统、AS-OCT设备和USP相比,CT-1P眼压角膜厚度仪低估了CCT测量值。USP、Pentacam和AS-OCT测量的平均CCT具有可比性,且存在显著的线性相关性。在临床实践中,这三种测量方式在健康患者中可相互替换。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验