使用光学低相干反射测量法、傅里叶域光学相干断层扫描术和Scheimpflug相机测量中央角膜厚度的比较。

Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera.

作者信息

Gonul Saban, Koktekir Bengu Ekinci, Bakbak Berker, Gedik Sansal

机构信息

Department of Ophthalmology, Selcuk University Faculty of Medicine, Konya, Turkey.

出版信息

Arq Bras Oftalmol. 2014 Nov-Dec;77(6):345-50. doi: 10.5935/0004-2749.20140087.

Abstract

PURPOSE

To compare the results of central corneal thickness (CCT) measurements obtained using optical low-coherence reflectometry (OLCR), Fourier domain optical coherence tomography (FD-OCT), and a Scheimpflug camera (SC), combined with Placido corneal topography.

METHODS

A total of 25 healthy subjects were enrolled in the present study, and one eye of each subject was included. A detailed ophthalmic examination was performed in all cases following CCT measurements with OLCR, FD-OCT, and SC. The results were compared using an ANOVA test. Bland-Altman analysis was used to demonstrate agreement between methods. Intra-examiner repeatability was assessed by using intraclass correlation coefficients (ICCs).

RESULTS

Statistically significant differences were observed between the results of the CCT measurements obtained using the three different devices (p=0.009). Significant correlations were found between OLCR and FD-OCT (r=0.97; p<0.0001), FD-OCT and SC (r=0.91; p<0.0001), and OLCR and SC (r=0.95; p<0.0001). The 95% limits of agreement (LOA) obtained from Bland-Altman plots were from -7.2 μm to 28.7 μm for OLCR versus FD-OCT, from -19.2 μm to 30.4 μm for OLCR versus SC, and from -42.6 to 32.3 μm for FD-OCT versus SC. Intra-examiner repeatability was excellent for each method, with ICCs >0.98.

CONCLUSION

Although the results of CCT measurements obtained from these three devices were highly correlated with one another and the mean differences between instruments were comparable with the reported diurnal CCT fluctuation, the measurements are not directly interchangeable in clinical practice because of the wide LOA values.

摘要

目的

比较使用光学低相干反射测量法(OLCR)、傅里叶域光学相干断层扫描(FD-OCT)以及与普拉西多角膜地形图相结合的Scheimpflug相机(SC)所获得的中央角膜厚度(CCT)测量结果。

方法

本研究共纳入25名健康受试者,每名受试者的一只眼睛被纳入研究。在使用OLCR、FD-OCT和SC测量CCT后,对所有病例进行了详细的眼科检查。使用方差分析检验比较结果。采用Bland-Altman分析来证明不同方法之间的一致性。通过组内相关系数(ICC)评估检查者内部的重复性。

结果

使用三种不同设备获得的CCT测量结果之间存在统计学显著差异(p = 0.009)。在OLCR与FD-OCT之间(r = 0.97;p < 0.0001)、FD-OCT与SC之间(r = 0.91;p < 0.0001)以及OLCR与SC之间(r = 0.95;p < 0.0001)发现了显著相关性。从Bland-Altman图获得的95%一致性界限(LOA),OLCR与FD-OCT之间为-7.2μm至28.7μm,OLCR与SC之间为-19.2μm至30.4μm,FD-OCT与SC之间为-42.6至32.3μm。每种方法的检查者内部重复性都非常好,ICC>0.98。

结论

尽管从这三种设备获得的CCT测量结果彼此高度相关,并且仪器之间的平均差异与报道的CCT昼夜波动相当,但由于LOA值较宽,这些测量结果在临床实践中不能直接互换。