High accuracy dosimetry with small pieces of Gafchromic films.

作者信息

Dąbrowski Ryszard, Drozdyk Izabela, Kukołowicz Paweł

机构信息

Medical Physics Department, Cancer Center-Institute of Oncology, Roentgena 5, 02-781, Warsaw, Poland.

出版信息

Rep Pract Oncol Radiother. 2018 Mar-Apr;23(2):114-120. doi: 10.1016/j.rpor.2018.01.001. Epub 2018 Feb 21.

Abstract

AIM

To investigate the influence of several factors on the accuracy of dose measurements and feasibility of application of small Gafchromic detectors for postal audit.

BACKGROUND

Our experience showed that precision of dose measurements with small pieces of Gafchromic films may be significantly improved.

MATERIALS AND METHODS

Gafchromic films with dimensions of 1 × 1, 2 × 2 and 3 × 3 cm were exposed to 6 MV X-rays at dose levels of 50 cGy-210 cGy. The single- and multichannel methods (MM) were used for dose measurements. Detectors were scanned with an Epson V750PRO flatbed colour scanner. For 1 × 1 and larger detector sizes, separate calibration curves were established. The influence of the following factors was investigated: the heterogeneity of Gafchromic detectors group for single- and MM, ambient thermal detector conditions, the dose delivered on the measurement accuracy, application of two separate calibration curves for the smallest and larger pieces of films.

RESULTS

The MM improves significantly the precision of dose measurement. The uncertainty attributed to detector active layer differences and scanner instabilities was about 1 cGy (1 StDev) regardless of dose and detector size. The ambient temperature of the environment in which films were stored after irradiation influenced the dose reading. Significant difference of transmission for detectors sized 1 × 1 and 2 × 2cm was observed. The maximal difference between applied dose and dose reading performed was 1.1%.

CONCLUSIONS

The MM with a scaling protocol leads to a very high precision of dose measurements. The ambient thermal detector environment causes significant changes of measured signal. The detector size has relevant impact on dose reading.

摘要

目的

研究多种因素对剂量测量准确性的影响以及小型Gafchromic探测器用于邮政审计的应用可行性。

背景

我们的经验表明,使用小块Gafchromic胶片进行剂量测量的精度可能会显著提高。

材料与方法

尺寸为1×1、2×2和3×3 cm的Gafchromic胶片在50 cGy - 210 cGy的剂量水平下接受6 MV X射线照射。采用单通道和多通道方法(MM)进行剂量测量。使用爱普生V750PRO平板彩色扫描仪对探测器进行扫描。对于1×1及更大尺寸的探测器,建立了单独的校准曲线。研究了以下因素的影响:单通道和多通道方法中Gafchromic探测器组的异质性、探测器的环境热条件、所给予剂量对测量准确性的影响、对最小和较大尺寸胶片应用两条单独校准曲线的情况。

结果

多通道方法显著提高了剂量测量的精度。无论剂量和探测器尺寸如何,探测器活性层差异和扫描仪不稳定性导致的不确定性约为1 cGy(1标准偏差)。照射后胶片储存环境的温度影响剂量读数。观察到尺寸为1×1和2×2 cm的探测器在透射率上存在显著差异。所施加剂量与测量剂量读数之间的最大差异为1.1%。

结论

采用缩放协议的多通道方法可实现非常高的剂量测量精度。探测器的环境热条件会导致测量信号发生显著变化。探测器尺寸对剂量读数有相关影响。

相似文献

1
High accuracy dosimetry with small pieces of Gafchromic films.
Rep Pract Oncol Radiother. 2018 Mar-Apr;23(2):114-120. doi: 10.1016/j.rpor.2018.01.001. Epub 2018 Feb 21.
3
Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry.
Phys Med Biol. 2009 Feb 21;54(4):1073-85. doi: 10.1088/0031-9155/54/4/017. Epub 2009 Jan 23.
4
Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance.
Phys Med Biol. 2007 Jul 21;52(14):4211-25. doi: 10.1088/0031-9155/52/14/013. Epub 2007 Jun 18.
5
Technical Note: Impact of region of interest size and location in Gafchromic film dosimetry.
Med Phys. 2018 May;45(5):2329-2336. doi: 10.1002/mp.12885. Epub 2018 Apr 18.
7
Multichannel film dosimetry with nonuniformity correction.
Med Phys. 2011 May;38(5):2523-34. doi: 10.1118/1.3576105.
9
Detector response in the buildup region of small MV fields.
Med Phys. 2020 Mar;47(3):1327-1339. doi: 10.1002/mp.13973. Epub 2020 Jan 20.
10
Characteristics of Gafchromic XRQA2 films for kV image dose measurement.
Med Phys. 2012 Feb;39(2):842-50. doi: 10.1118/1.3675398.

引用本文的文献

1
In-vivo dosimetry with Gafchromic films for patients with sarcoma.
Rep Pract Oncol Radiother. 2025 Jun 7;30(2):129-137. doi: 10.5603/rpor.104982. eCollection 2025.

本文引用的文献

1
A method for multichannel dosimetry with EBT3 radiochromic films.
Med Phys. 2014 Jun;41(6):062101. doi: 10.1118/1.4871622.
3
Enhanced dosimetry procedures and assessment for EBT2 radiochromic film.
Med Phys. 2012 Apr;39(4):2147-55. doi: 10.1118/1.3694100.
4
Multichannel film dosimetry with nonuniformity correction.
Med Phys. 2011 May;38(5):2523-34. doi: 10.1118/1.3576105.
8
Precise radiochromic film dosimetry using a flat-bed document scanner.
Med Phys. 2005 Jul;32(7):2245-53. doi: 10.1118/1.1929253.
9
Tolerances for the accuracy of photon beam dose calculations of treatment planning systems.
Radiother Oncol. 2001 Aug;60(2):191-201. doi: 10.1016/s0167-8140(01)00377-2.
10
The IAEA/WHO TLD postal programme for radiotherapy hospitals.
Radiother Oncol. 2000 Jan;54(1):65-72. doi: 10.1016/s0167-8140(99)00164-4.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验