Suppr超能文献

皮下植入式心脏复律除颤器疗效的决定因素:计算机建模研究。

Determinants of Subcutaneous Implantable Cardioverter-Defibrillator Efficacy: A Computer Modeling Study.

机构信息

Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, Massachusetts.

Independent Researcher, Minneapolis, Minnesota.

出版信息

JACC Clin Electrophysiol. 2017 Apr;3(4):405-414. doi: 10.1016/j.jacep.2016.10.016. Epub 2017 Mar 1.

Abstract

OBJECTIVES

This study determined the impact of subcutaneous implantable cardioverter-defibrillator (S-ICD) coil and generator position on defibrillation threshold (DFT).

BACKGROUND

S-ICD implantation can occasionally result in unacceptably high DFT. Implant position characteristics associated with high DFTs in S-ICD patients have not been fully elucidated.

METHODS

A 3.8-million-element computer model built from magnetic resonance images was used to simulate the electric fields that occur during defibrillation. Generator positions were tested from posterior to anterior in 4-cm increments. The left parasternal coil was tested with 0, 5, and 10 mm of underlying subcutaneous fat and the generator with 20 mm of underlying fat. The estimated DFT for the S-ICD was defined as the energy delivered when producing an electric field of 4 volts/cm in at least 95% of the ventricular myocardium.

RESULTS

Estimated DFTs were 22, 29, 64, and 135 joules for posterior, standard (lateral), mid-anterior, and anterior generator locations, respectively. Defibrillation thresholds were 29, 58, and 95 joules with 0, 5, and 10 mm subcoil fat, respectively, and 45 joules with 20 mm subgenerator fat. Combining anterior generator position with subcoil fat resulted in a very high DFT (379 joules). Shock impedance increased with both subcoil and subgenerator fat but was minimally affected by anterior/posterior generator position.

CONCLUSIONS

The model suggests that an S-ICD implantation strategy involving posterior generator location and coil and generator directly over the fascia without underlying fat is likely to markedly lower DFTs with the S-ICD and assist in troubleshooting of patients with unacceptably high DFTs.

摘要

目的

本研究旨在确定皮下植入式心律转复除颤器(S-ICD)的线圈和发生器位置对除颤阈值(DFT)的影响。

背景

S-ICD 的植入偶尔会导致无法接受的高 DFT。与 S-ICD 患者高 DFT 相关的植入位置特征尚未完全阐明。

方法

使用基于磁共振成像构建的 380 万个元素的计算机模型来模拟除颤过程中发生的电场。以 4 厘米为增量,从前向后测试发生器位置。左胸骨旁线圈分别测试了无、5 毫米和 10 毫米的皮下脂肪以及 20 毫米的皮下脂肪下的发生器。S-ICD 的估计 DFT 定义为在至少 95%的心室心肌中产生 4 伏/厘米电场时输送的能量。

结果

分别为后、标准(侧)、中前和前发生器位置的估计 DFT 为 22、29、64 和 135 焦耳。无、5 毫米和 10 毫米皮下线圈脂肪的 DFT 分别为 29、58 和 95 焦耳,20 毫米皮下发生器脂肪的 DFT 为 45 焦耳。在前部发生器位置和皮下线圈脂肪的结合导致非常高的 DFT(379 焦耳)。随着皮下线圈和发生器脂肪的增加,电击阻抗增加,但受前/后发生器位置的影响最小。

结论

该模型表明,涉及后部发生器位置以及线圈和发生器直接位于筋膜上而无皮下脂肪的 S-ICD 植入策略可能会显著降低 S-ICD 的 DFT,并有助于解决 DFT 不可接受的患者的故障排除。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验