你能低到什么程度?五种自动化睾酮免疫测定法的分析性能。

How low can you go? Analytical performance of five automated testosterone immunoassays.

作者信息

La'ulu Sonia L, Kalp Kimberly J, Straseski Joely A

机构信息

ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 84108, USA.

Department of Pathology, University of Utah Health Sciences Center, Salt Lake City, UT 84108, USA.

出版信息

Clin Biochem. 2018 Aug;58:64-71. doi: 10.1016/j.clinbiochem.2018.05.008. Epub 2018 May 12.

Abstract

BACKGROUND

Testosterone is commonly measured using immunoassays, yet concerns with the accuracy and quality of testing by these methods exist, particularly for low testosterone concentrations. Study objectives were to evaluate selective performance characteristics, including functional sensitivity (FS), of 5 automated immunoassays for total testosterone.

METHODS

FS, imprecision, assay interference, limit of blank, linearity, and accuracy were assessed using the Abbott ARCHITECT i2000, SIEMENS ADVIA Centaur and IMMULITE 2000, Beckman Coulter DxI 800, and Roche MODULAR E170. Comparisons to an in-house liquid chromatography-tandem mass spectrometry (LC-MS/MS) method were performed using patient samples from men, women, boys, and girls.

RESULTS

FS at 20% coefficient of variation (CV) for the ARCHITECT, Centaur, DxI, E170 and IMMULITE assays were 0.14, 1.23, 0.36, 0.77, 3.49 nmol/L, respectively. Total CVs for the 5-day imprecision study were ≤ 9.0% for all methods. All assays met manufacturer's claims for hemolysis, icterus, and lipemia interference and limit of blank. Dilution linearity studies had deviations from the target recoveries ranging from 3.4% (ARCHITECT) to 14.3% (DxI). Using National Institute of Standards and Technology Standard Reference Material 971, recoveries ranged from 79.2-149.2% (DxI, male and female, respectively). When compared to LC-MS/MS, more immunoassays under-recovered in men and women and over-recovered in boys and girls. Slopes ranged from 0.71 (IMMULITE, women) to 1.35 (DxI, boys). The combined average for percent bias was higher in boys (28.0%) than men (11.6%), women (22.8%), and girls (25.7%).

CONCLUSIONS

Challenges with accurately measuring testosterone appear to remain for some immunoassays, but not all. While most immunoassays remain optimized for concentrations observed in healthy men, some showed acceptable performance when challenged at lower concentrations.

摘要

背景

睾酮通常采用免疫测定法进行检测,但人们对这些方法检测的准确性和质量存在担忧,尤其是对于低睾酮浓度的情况。研究目的是评估5种总睾酮自动化免疫测定法的选择性性能特征,包括功能灵敏度(FS)。

方法

使用雅培ARCHITECT i2000、西门子ADVIA Centaur和IMMULITE 2000、贝克曼库尔特DxI 800以及罗氏MODULAR E170评估FS、不精密度、分析干扰、空白限、线性和准确性。使用来自男性、女性、男孩和女孩的患者样本与内部液相色谱 - 串联质谱(LC - MS/MS)方法进行比较。

结果

ARCHITECT、Centaur、DxI、E170和IMMULITE测定法在变异系数(CV)为20%时的FS分别为0.14、1.23、0.36、0.77、3.49 nmol/L。所有方法的5天不精密度研究的总CV均≤9.0%。所有测定法均符合制造商关于溶血、黄疸和脂血干扰以及空白限的要求。稀释线性研究的回收率与目标回收率的偏差范围为3.4%(ARCHITECT)至14.3%(DxI)。使用美国国家标准与技术研究院标准参考物质971,回收率范围为79.2 - 149.2%(分别为DxI的男性和女性样本)。与LC - MS/MS相比,更多免疫测定法在男性和女性中回收不足,而在男孩和女孩中回收过量。斜率范围为0.71(IMMULITE,女性)至1.35(DxI,男孩)。偏差百分比的综合平均值在男孩中(28.0%)高于男性(11.6%)、女性(22.8%)和女孩(25.7%)。

结论

对于某些免疫测定法而言,准确测量睾酮似乎仍然存在挑战,但并非所有测定法都如此。虽然大多数免疫测定法仍针对健康男性中观察到的浓度进行了优化,但有些在较低浓度下接受挑战时表现出可接受的性能。