La'ulu Sonia L, Rawlins Mindy L, Pfeiffer Christine M, Zhang Mindy, Roberts William L
ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 84108, USA.
Am J Clin Pathol. 2008 Dec;130(6):969-75. doi: 10.1309/AJCP64BJIPNPSQDJ.
Elevated concentrations of homocysteine (Hcy) are associated with a range of disorders. Linearity, imprecision, interference, method comparison, and accuracy were evaluated on the ADVIA Centaur (Siemens Healthcare Diagnostics, Deerfield, IL), ARCHITECT i2000SR (Abbott Diagnostics, Abbott Park, IL), AxSYM (Abbott Diagnostics), and IMMULITE 2000 (Siemens Healthcare Diagnostics) methods and analyzers and the Catch (Equal Diagnostics, Exton, PA) and Diazyme (Diazyme Laboratories, San Diego, CA) methods, both on the Modular P analyzer (Roche Diagnostics, Indianapolis, IN). All methods were linear with maximum deviations from target recoveries of less than 10%. Total coefficients of variation ranged from 1.7% to 9.4%. The effects of hemolysis, icterus, and lipemia were assessed. Method comparisons were performed using high-performance liquid chromatography as the comparison method. Correlation coefficients were 0.95 to 0.99. Bland-Altman plots demonstrated percentage bias of -29.3% (IMMULITE) to 7.2% (Centaur). Accuracy using the National Institute of Standards and Technology Standard Reference Material 1955 showed varying results with only 1 method within the certified range for all 3 levels. All methods demonstrated acceptable performance except the IMMULITE, which is less precise and accurate. Standardization of most methods seems acceptable, although continuing efforts are warranted.
高同型半胱氨酸(Hcy)浓度与一系列疾病相关。对ADVIA Centaur(西门子医疗诊断公司,伊利诺伊州迪尔菲尔德)、ARCHITECT i2000SR(雅培诊断公司,伊利诺伊州雅培公园)、AxSYM(雅培诊断公司)和IMMULITE 2000(西门子医疗诊断公司)的方法及分析仪,以及在Modular P分析仪(罗氏诊断公司,印第安纳州印第安纳波利斯)上的Catch(Equal诊断公司,宾夕法尼亚州埃克斯顿)和Diazyme(Diazyme实验室,加利福尼亚州圣地亚哥)方法,进行了线性、不精密度、干扰、方法比较和准确性评估。所有方法均呈线性,与目标回收率的最大偏差小于10%。总变异系数范围为1.7%至9.4%。评估了溶血、黄疸和脂血的影响。使用高效液相色谱法作为比较方法进行方法比较。相关系数为0.95至0.99。Bland-Altman图显示偏差百分比为-29.3%(IMMULITE)至7.2%(Centaur)。使用美国国家标准与技术研究院标准参考物质1955的准确性显示结果各异,所有3个水平中只有1种方法在认证范围内。除IMMULITE外,所有方法均表现出可接受的性能,IMMULITE的精密度和准确性较低。大多数方法的标准化似乎可以接受,尽管仍需持续努力。