Büther Florian, Noto Benjamin, Auf der Springe Katharina, Allkemper Thomas, Stegger Lars
1Department of Nuclear Medicine, University Hospital Münster, Albert-Schweitzer-Campus 1, 48149 Münster, Germany.
2European Institute for Molecular Imaging, University of Münster, Münster, Germany.
Eur J Hybrid Imaging. 2017;1(1):10. doi: 10.1186/s41824-017-0015-x. Epub 2017 Nov 8.
Attenuation correction is one of the most important steps in producing quantitative PET image data. In hybrid PET-MRI systems, this correction is far from trivial, as MRI data are not correlated to PET attenuation properties of the scanned object. Commercially available systems often employ correction schemes based on segmenting the body into different tissue classes (air, lung tissue, fat-, and water-like soft tissue), e.g. by using a dual time-point Dixon sequence. However, several pitfalls are known for this approach. Here a specific artefact of MR-based PET attenuation correction is reported, caused by misidentifying the liver as lung tissue due to iron overload.
A patient with a history of hematopoietic stem cell transplantation underwent a whole-body [F]FDG PET-MRI scan. Markedly low liver uptake values were noted in the PET images, seemingly caused by an erroneous assignment of lung tissue attenuation values to the liver. A closer investigation demonstrated markedly low MRI intensity values of the liver, indicative of secondary hemochromatosis (iron overload) most probably due to a history of multiple blood transfusions. Manual assignment of adequate liver attenuation values resulted in more realistic PET images.
Iron overload of the liver was identified as a cause of a specific attenuation correction artefact. It remains to be seen how frequent this artefact will be encountered; however, this case highlights that attenuation maps should always be checked during PET image interpretation in hybrid PET-MRI.
衰减校正是生成定量PET图像数据最重要的步骤之一。在PET-MRI混合系统中,这种校正绝非易事,因为MRI数据与被扫描物体的PET衰减特性不相关。市售系统通常采用基于将身体分割为不同组织类别(空气、肺组织、脂肪样和水样软组织)的校正方案,例如通过使用双时间点狄克逊序列。然而,这种方法存在一些缺陷。本文报道了一种基于MR的PET衰减校正的特定伪影,该伪影是由于铁过载将肝脏误识别为肺组织所致。
一名有造血干细胞移植病史的患者接受了全身[F]FDG PET-MRI扫描。PET图像中肝脏摄取值明显偏低,似乎是由于将肺组织衰减值错误地指定给了肝脏。进一步检查显示肝脏的MRI强度值明显偏低,这表明最有可能由于多次输血史导致继发性血色素沉着症(铁过载)。手动指定适当的肝脏衰减值可得到更真实的PET图像。
肝脏铁过载被确定为一种特定衰减校正伪影的原因。这种伪影的出现频率还有待观察;然而,该病例强调在PET-MRI混合系统中进行PET图像解读时应始终检查衰减图。