Department of Neurological Surgery, Rush University Medical Center, Chicago, Illinois, USA.
Department of Neurosurgery, Wakayama Medical University, Wakayama City, Japan.
J Neurointerv Surg. 2019 Jan;11(1):74-79. doi: 10.1136/neurintsurg-2018-013935. Epub 2018 May 26.
To evaluate the rate of neointimal development and thrombus formation of surface-modified flow diverters in single antiplatelet therapy (SAPT) using optical coherence tomography (OCT) in a porcine model.
We divided 10 experimental pigs into two groups. One group (n=6) received dual antiplatelet therapy (DAPT) and the other group (n=4) received SAPT. Four stents (two per carotid artery) were implanted in both groups. The stents used were the Pipeline Flex embolization device (PED Flex), Pipeline Flex with Shield technology (PED Shield), and the Solitaire AB stent. All animals underwent weekly angiography and OCT. The OCT data were analyzed using the following measurements: neointimal ratio ((stent - lumen area)/stent area), stent-coverage ratio (number of stent struts covered by neointima/total stent struts), and the presence or absence of thrombus formation per 1 mm cross-section.
PED Flex and Shield in the SAPT group had higher neointimal ratios than in the DAPT group (P<0.001, respectively). In the DAPT group, the speed of endothelial growth on day 7 in the PED Shield group was higher than that in the PED Flex group (P<0.001). In the SAPT group, PED Flex demonstrated significantly more thrombus formation on day 7 than PED Shield (P<0.001).
The PED Shield stent showed faster endothelial growth than the other devices and comparable neointimal volume. There was significantly less thrombus formation on PED Shield than PED Flex when using SAPT in a porcine model.
通过光学相干断层扫描(OCT)评估猪模型中单抗血小板治疗(SAPT)中表面改性的血流导向装置的新生内膜形成和血栓形成率。
我们将 10 只实验猪分为两组。一组(n=6)接受双联抗血小板治疗(DAPT),另一组(n=4)接受 SAPT。两组均植入四个支架(每个颈动脉两个)。使用的支架为 Pipeline Flex 栓塞装置(PED Flex)、带有 Shield 技术的 Pipeline Flex(PED Shield)和 Solitaire AB 支架。所有动物均进行每周血管造影和 OCT 检查。使用以下测量值分析 OCT 数据:新生内膜比((支架-管腔面积)/支架面积)、支架覆盖率(被新生内膜覆盖的支架支架数/总支架支架数)以及每 1mm 横截面存在或不存在血栓形成。
SAPT 组的 PED Flex 和 Shield 的新生内膜比高于 DAPT 组(分别为 P<0.001)。在 DAPT 组中,PED Shield 组第 7 天内皮生长速度高于 PED Flex 组(P<0.001)。在 SAPT 组中,PED Flex 在第 7 天的血栓形成明显多于 PED Shield(P<0.001)。
PED Shield 支架的内皮生长速度比其他装置快,新生内膜体积相当。在猪模型中,SAPT 时 PED Shield 上血栓形成明显少于 PED Flex。