Correlation between Insertion Torque and Implant Stability Quotient in Tapered Implants with Knife-Edge Thread Design.

机构信息

Department of Surgical Sciences, University of Genoa, Italy.

Private Practice, Cassano allo Ionio, Italy.

出版信息

Biomed Res Int. 2018 May 15;2018:7201093. doi: 10.1155/2018/7201093. eCollection 2018.

Abstract

AIM

To evaluate the correlation between insertion torque (IT) and implant stability quotient (ISQ) in tapered implants with knife-edge threads.

METHODS

Seventy-five identical implants (Anyridge, Megagen) were inserted by using a surgical drilling unit with torque control and an integrated resonance frequency analysis module (Implantmed, W&H). IT (N/cm) and ISQ were recorded and implants were divided into three groups ( = 25) according to the IT: low (<30), medium (30 < IT < 50), and high torque (>50). ISQ difference among groups was assessed by Kruskal-Wallis test, followed by Bonferroni-corrected Mann-Whitney -test for pairwise comparisons. The strength of the association between IT and ISQ was assessed by Spearman Rho correlation coefficient (α = 0.05).

RESULTS

At the pairwise comparisons, a significant difference of ISQ values was demonstrated only between low torque and high torque groups. The strength of the association between IT and ISQ value was significant for both the entire sample and the medium torque group, while it was not significant in low and high torque groups.

CONCLUSIONS

For the investigated implant, ISQ and IT showed a positive correlation up to values around 50 N/cm: higher torques subject the bone-implant system to unnecessary biological and mechanical stress without additional benefits in terms of implant stability. This trial is registered with NCT03222219.

摘要

目的

评估刃状螺纹锥形种植体的植入扭矩(IT)与种植体稳定性系数(ISQ)之间的相关性。

方法

使用带有扭矩控制和集成共振频率分析模块的外科钻孔装置(Implantmed,W&H)插入 75 个相同的种植体(Anyridge,Megagen)。记录 IT(N/cm)和 ISQ,并根据 IT 将种植体分为三组(n=25):低(<30)、中(30 < IT < 50)和高扭矩(>50)。采用 Kruskal-Wallis 检验评估组间 ISQ 差异,然后采用 Bonferroni 校正的 Mann-Whitney U 检验进行两两比较。采用 Spearman Rho 相关系数(α=0.05)评估 IT 和 ISQ 值之间的关联强度。

结果

在两两比较中,仅在低扭矩和高扭矩组之间观察到 ISQ 值的显著差异。IT 和 ISQ 值之间的关联强度在整个样本和中扭矩组中均具有统计学意义,而在低扭矩和高扭矩组中则无统计学意义。

结论

对于所研究的种植体,ISQ 和 IT 之间呈正相关,直至约 50 N/cm:更高的扭矩会使骨-种植体系统承受不必要的生物和机械应力,而在种植体稳定性方面没有额外的益处。本试验已在 NCT03222219 注册。