便携式间接测热仪测量癌症患者静息能量消耗的准确性。

Accuracy of a Portable Indirect Calorimeter for Measuring Resting Energy Expenditure in Individuals With Cancer.

机构信息

Human Nutrition Research Unit, Department of Agricultural, Food, and Nutritional Science, Faculty of Agricultural, Life, and Environmental Sciences, University of Alberta, Edmonton, Canada.

Alberta Research Centre for Health Evidence, Department of Pediatrics, University of Alberta, Edmonton, Canada.

出版信息

JPEN J Parenter Enteral Nutr. 2019 Jan;43(1):145-151. doi: 10.1002/jpen.1310. Epub 2018 Jun 5.

Abstract

BACKGROUND

Determining optimal caloric intake for an individual with cancer is complicated by metabolic changes that occur, namely, alterations in resting energy expenditure (REE). There is currently no validated clinically available equation or tool to measure energy expenditure in these patients.

METHODS

Patients with newly diagnosed solid tumors underwent REE assessments using the FitMate GS portable indirect calorimeter and reference VMax metabolic cart; both used canopy hoods. REE was also estimated from the Harris-Benedict, Mifflin St. Jeor, and Henry equations for comparison. Data were analyzed using paired samples t-test and the Bland-Altman approach to assess group-level and individual-level agreement compared with the metabolic cart.

RESULTS

A total 26 patients (19 males; body mass index: 27.8 ± 5.5 kg/m ; age: 62 ± 10 years) participated in the study. Biases for the FitMate GS and both equations were low (ranging from -44 to -92 kcal or -2.3% to -5.1%), indicating good group-level accuracy. The FitMate GS had low bias, but the widest limits of agreement (-28.0% to 21.2%) compared with the 3 equations (Harris-Benedict: -15.8% to 11.2%; Mifflin St. Jeor: -17.1% to 6.9%; Henry: -15.4% to 11.5%). These differences were not due to volume of oxygen, BMI category, or sex.

CONCLUSION

FitMate GS performed well on a group level, but its accuracy was poor on an individual level. Further research should develop better equations and validate tools to measure energy expenditure for accurate dietary recommendations for patients at nutrition risk.

摘要

背景

癌症患者的最佳热量摄入受到代谢变化的影响,即静息能量消耗 (REE) 的改变。目前,没有经过验证的临床可用方程或工具可以测量这些患者的能量消耗。

方法

新诊断为实体瘤的患者使用 FitMate GS 便携式间接测热仪和参考 VMax 代谢箱进行 REE 评估;两种仪器均使用罩篷罩。REE 还可以通过 Harris-Benedict、Mifflin St. Jeor 和 Henry 方程进行估算,以进行比较。使用配对样本 t 检验和 Bland-Altman 方法分析数据,以评估与代谢箱相比的组内和个体内的一致性。

结果

共有 26 名患者(19 名男性;BMI:27.8 ± 5.5 kg/m ;年龄:62 ± 10 岁)参与了研究。FitMate GS 和两个方程的偏差都较低(范围为-44 至-92 kcal 或-2.3%至-5.1%),表明组内准确性较好。FitMate GS 的偏差较低,但与三个方程(Harris-Benedict:-15.8%至 11.2%;Mifflin St. Jeor:-17.1%至 6.9%;Henry:-15.4%至 11.5%)相比,其一致性范围最宽(-28.0%至 21.2%)。这些差异不是由于氧气量、BMI 类别或性别引起的。

结论

FitMate GS 在组内水平上表现良好,但在个体水平上准确性较差。进一步的研究应开发更好的方程和验证工具,以测量能量消耗,为营养风险患者提供准确的饮食建议。