Thiebaud Robert S, Funk Merrill D, Patton Jacelyn C, Massey Brook L, Shay Terri E, Schmidt Martin G, Giovannitti Nicolas
1Department of Kinesiology, Texas Wesleyan University, USA.
2Department of Health and Human Performance, University of Texas Rio Grande Valley, USA.
Digit Health. 2018 Apr 13;4:2055207618770322. doi: 10.1177/2055207618770322. eCollection 2018 Jan-Dec.
The ability to monitor physical activity throughout the day and during various activities continues to improve with the development of wrist-worn monitors. However, the accuracy of wrist-worn monitors to measure both heart rate and energy expenditure during physical activity is still unclear. The purpose of this study was to determine the accuracy of several popular wrist-worn monitors at measuring heart rate and energy expenditure.
Participants wore the TomTom Cardio, Microsoft Band and Fitbit Surge on randomly assigned locations on each wrist. The maximum number of monitors per wrist was two. The criteria used for heart rate and energy expenditure were a three-lead electrocardiogram and indirect calorimetry using a metabolic cart. Participants exercised on a treadmill at 3.2, 4.8, 6.4, 8 and 9.7 km/h for 3 minutes at each speed, with no rest between speeds. Heart rate and energy expenditure were manually recorded every minute throughout the protocol.
Mean absolute percentage error for heart rate varied from 2.17 to 8.06% for the Fitbit Surge, from 1.01 to 7.49% for the TomTom Cardio and from 1.31 to 7.37% for the Microsoft Band. The mean absolute percentage error for energy expenditure varied from 25.4 to 61.8% for the Fitbit Surge, from 0.4 to 26.6% for the TomTom Cardio and from 1.8 to 9.4% for the Microsoft Band.
Data from these devices may be useful in obtaining an estimate of heart rate for everyday activities and general exercise, but energy expenditure from these devices may be significantly over- or underestimated.
随着腕部佩戴式监测器的发展,全天及各类活动期间监测身体活动的能力不断提高。然而,腕部佩戴式监测器在身体活动期间测量心率和能量消耗的准确性仍不明确。本研究的目的是确定几款流行的腕部佩戴式监测器在测量心率和能量消耗方面的准确性。
参与者在每只手腕的随机指定位置佩戴TomTom Cardio、Microsoft Band和Fitbit Surge。每只手腕佩戴监测器的最大数量为两个。用于心率和能量消耗的标准是三导联心电图和使用代谢车进行间接量热法。参与者在跑步机上以3.2、4.8、6.4、8和9.7公里/小时的速度各运动3分钟,速度之间不休息。在整个实验过程中每分钟手动记录心率和能量消耗。
Fitbit Surge的心率平均绝对百分比误差在2.17%至8.06%之间,TomTom Cardio在1.01%至7.49%之间,Microsoft Band在1.31%至7.37%之间。Fitbit Surge的能量消耗平均绝对百分比误差在25.4%至61.8%之间,TomTom Cardio在0.4%至26.6%之间,Microsoft Band在1.8%至9.4%之间。
这些设备的数据可能有助于获取日常活动和一般锻炼的心率估计值,但这些设备的能量消耗可能会被显著高估或低估。