IEEE Trans Biomed Eng. 2019 Jan;66(1):4-13. doi: 10.1109/TBME.2018.2834361. Epub 2018 May 8.
Ventral hernia repairs using mesh prosthetics suffer from high recurrence rates, with 10%-20% of repairs failing within three years. Uneven distribution of stress within the implanted mesh prosthetic is thought to contribute to the high recurrence rate. We propose a method for providing quantitative guidance and monitoring of hernia repairs using an array of magnetoelastic strain sensors.
The magnetoelastic strain sensors presented here are based on a coupled design to achieve measurements with higher signal-to-noise ratio (SNR). A first magnetoelastic element (the transducer) is bonded to the mesh prosthetic and is characterized by a strain-dependent magnetic field. The resonance frequency of a second magnetoelastic element (the resonator) encased in a rigid casing is biased by the transducer element's magneticity and can be measured noninvasively using an external interrogation coil. The coupled magnetoelastic strain sensors are assembled using a combination of photochemical machining, patterning, and heat sealing.
The dynamic range of the coupled sensors can be tuned by altering the transducer geometry. Additional spring elements are integrated onto the transducer element to achieve high dynamic range measurements saturating at 74 millistrains.
A coupled magnetoelastic strain sensor combines a transducer with an encased resonator element to measure strain with high SNR on an implantable flexible hernia mesh substrate.
This study provides surgeons and researchers with a clinically relevant tool to quantify the strain distributions within implanted mesh prosthetics, with the ultimate goal of reducing the recurrence rate of ventral hernia repairs.
使用网片修复腹疝的复发率很高,有 10%-20%的修复在三年内失败。植入网片内的应力分布不均匀被认为是导致高复发率的原因之一。我们提出了一种使用磁弹性应变传感器阵列为疝修补术提供定量指导和监测的方法。
这里提出的磁弹性应变传感器基于一种耦合设计,以实现具有更高信噪比 (SNR) 的测量。第一个磁弹性元件(换能器)粘接到网片上,并具有应变相关的磁场。封装在刚性外壳中的第二个磁弹性元件(谐振器)的谐振频率由换能器元件的磁性偏置,可以使用外部询问线圈进行非侵入式测量。耦合磁弹性应变传感器使用光化学加工、图案化和热密封的组合进行组装。
通过改变换能器的几何形状可以调整耦合传感器的动态范围。在换能器元件上集成了附加的弹簧元件,以实现高达 74 毫应变的高动态范围测量。
耦合磁弹性应变传感器将换能器与封装的谐振器元件结合在一起,在可植入的柔性疝网片基底上以高 SNR 测量应变。
这项研究为外科医生和研究人员提供了一种临床相关的工具,可量化植入网片内的应变分布,最终目标是降低腹疝修复的复发率。