Ultrasound guided needle biopsy is an important method for collection of breast cancer tissue. In this paper, we report on the design and testing of a high-voltage 1 to 64 Multiplexer/Demultiplexer (MUX/De-MUX) integrated circuit (IC) for ultrasound-guided breast biopsy applications implemented in a high-voltage CMOS process. The IC is intended to be incorporated inside the breast biopsy needle and is designed to fit inside the needle inner diameter of 2.38 mm. The MUX/De-MUX electronics are made up of three parts, including a low-voltage 6 to 64 decoder, a level shifter to convert from low voltage to high voltage, and analog high-voltage switches. Experimental results show a -3-dB bandwidth of over 70 MHz, Rds (on) of , -2.279-dB insertion loss, and -17.5-dB off isolation at 70 MHz with low-voltage input. Finally, we present results obtained via synthetic aperture imaging using the fabricated MUX/De-Mux device and a high-frequency ultrasound array. This device and technique hold promise for high-frequency imaging probes where a limited number of elements are used and the depth of penetration is short such as in breast biopsy and intravascular applications.
超声引导下的针吸活检是获取乳腺癌组织的重要方法。本文介绍了一种用于超声引导下乳腺活检应用的高压 1 至 64 复用器/解复用器(MUX/De-MUX)集成电路(IC)的设计和测试,该 IC 采用高压 CMOS 工艺实现。该 IC 旨在集成在乳腺活检针内,设计适合 2.38mm 内径的活检针。MUX/De-MUX 电子设备由三个部分组成,包括低压 6 至 64 解码器、用于从低压转换为高压的电平转换器以及模拟高压开关。实验结果表明,在 70MHz 时,具有低压输入的 3dB 带宽超过 70MHz,Rds(on)为 ,-2.279dB 的插入损耗和-17.5dB 的隔离度。最后,我们展示了使用制造的 MUX/De-Mux 器件和高频超声阵列进行合成孔径成像获得的结果。该器件和技术有望应用于高频成像探头,其中使用的元件数量有限,穿透深度较浅,如乳腺活检和血管内应用。