Suppr超能文献

用于腰椎成像的原型三维锥形束计算机断层摄影系统的初步评估,评估直立位置的人体尸体标本。

Initial Assessment of a Prototype 3D Cone-Beam Computed Tomography System for Imaging of the Lumbar Spine, Evaluating Human Cadaveric Specimens in the Upright Position.

机构信息

From the Clinic of Radiology and Nuclear Medicine, University Hospital Basel, University of Basel.

Amsler Consulting, Basel, Switzerland.

出版信息

Invest Radiol. 2018 Dec;53(12):714-719. doi: 10.1097/RLI.0000000000000495.

Abstract

OBJECTIVES

The aims of this study were to assess feasibility, image quality, and radiation dose and to estimate the optimal dose protocol for the lumbar spine of cadaveric specimens with different body mass indices (BMIs) in the upright position using a prototype 3-dimensional cone-beam computed tomography (CT) software implemented on a robotic x-ray system and compare with CT.

MATERIALS AND METHODS

The lumbar spine of 5 formalin-fixed human cadaveric specimens (BMI, 22-35 kg/m) was prospectively assessed in the upright position using prototype software for 3-dimensional tomography implemented on a robotic x-ray system. Specimens were scanned with varying kilovolt values (70, 81, 90, 100, 109, 121 kV) and thereafter with 80 kV (BMI ≤30 kg/m) and 121 kV (BMI >30 kg/m) and varying dose levels (DLs; 0.278, 0.435, 0.548, 0.696, 0.87, 1.09). Computed tomography data were acquired with a standard clinical protocol. Two independent readers rated visibility of the cortex, endplates, facet joints, trabeculae, neuroforamina, posterior alignment, and spinal canal as well as nerve roots. Radiation dose was measured with a cylindrical CTDI phantom. Descriptive statistics and analysis of variance were used (P < 0.05).

RESULTS

Average intraclass correlation was excellent (0.94). The lowest technically possible kilovolt and the highest technically possible DL yielded the best image quality. In specimens with a BMI of 30 kg/m or less, depiction of all structures was good and comparable to CT, except for nerve roots. For specimens with a BMI greater than 30 kg/m, image quality was limited.

CONCLUSIONS

Three-dimensional cone-beam CT of the lumbar spine in cadaveric specimens in the upright position is feasible. An optimal dose protocol was estimated. Depiction of osseous structures is comparable to CT in specimens with BMI of 30 kg/m or less. Image quality is limited for soft tissue structures and specimens with BMI greater than 30 kg/m.

摘要

目的

本研究旨在评估可行性、图像质量和辐射剂量,并估算在机器人 X 射线系统上使用原型三维锥形束 CT(CT)软件对不同体重指数(BMI)的直立尸体腰椎进行成像的最佳剂量方案,并与 CT 进行比较。

材料和方法

前瞻性评估 5 具福尔马林固定人体尸体标本(BMI 22-35kg/m)在直立位置时使用机器人 X 射线系统上实现的原型三维断层扫描软件。标本分别以 70、81、90、100、109、121kV 的不同千伏值进行扫描,然后以 80kV(BMI≤30kg/m)和 121kV(BMI>30kg/m)进行扫描,并以不同剂量水平(0.278、0.435、0.548、0.696、0.87、1.09)进行扫描。CT 数据采用标准临床方案采集。两名独立的读者评估皮质、终板、关节突关节、小梁、神经孔、后向排列和椎管以及神经根的可见度。使用圆柱形 CTDI 体模测量辐射剂量。采用描述性统计和方差分析(P<0.05)。

结果

平均组内相关系数为优秀(0.94)。最低技术可行千伏和最高技术可行剂量水平产生最佳的图像质量。BMI 为 30kg/m 或更低的标本中,除神经根外,所有结构的显示均良好,与 CT 相当。BMI 大于 30kg/m 的标本中,图像质量受到限制。

结论

在直立尸体标本中使用机器人 X 射线系统进行三维锥形束 CT 是可行的。估算出最佳剂量方案。BMI 为 30kg/m 或更低的标本中,骨结构的显示与 CT 相当。对于软组织结构和 BMI 大于 30kg/m 的标本,图像质量受到限制。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验