一种新型超快 T2 加权序列在脊柱成像中的临床应用。

Clinical Utility of a Novel Ultrafast T2-Weighted Sequence for Spine Imaging.

机构信息

From the Departments of Electrical and Computer Engineering (M.B.K., A.B.).

Medical Imaging (M.B.K., B.W., J.L.B., M.I.A., M.S.).

出版信息

AJNR Am J Neuroradiol. 2018 Aug;39(8):1568-1575. doi: 10.3174/ajnr.A5713. Epub 2018 Jul 12.

Abstract

BACKGROUND AND PURPOSE

TSE-based T2-weighted imaging of the spine has long scan times. This work proposes a fast imaging protocol using variable refocusing flip angles, optimized for blurring and specific absorption rate.

MATERIALS AND METHODS

A variable refocusing flip angle echo-train was optimized for the spine to improve the point spread function and minimize the specific absorption rate, yielding images with improved spatial resolution and SNR compared with the constant flip angle sequence. Data were acquired from 51 patients (35 lumbar, 16 whole-spine) using conventional TSE and the proposed sequence, with a single-shot variant for whole-spine. Noninferiority analysis was performed to evaluate the efficiency of the proposed technique.

RESULTS

The proposed multishot sequence resulted in a 2× shorter scan time with a >1.5× lower specific absorption rate. The variable flip angle sequence was noninferior to the conventional TSE ( < .025) for all image-quality and clinical criteria except signal-to-noise ratio for the lumbar spine protocol. However, mean image scores for the TSE-variable refocusing flip angle were ≥4.3 for all criteria, and concordance analysis showed high agreement (>90%) with the TSE, indicating clinical equivalence. The single-shot sequence resulted in 4× shorter whole-spine scans, and image scores were ≥4.4 for all criteria, attesting to its clinical utility.

CONCLUSIONS

We present a fast T2-weighted spine protocol using variable refocusing flip angles, including a single-shot variant. The sequences have better point spread function behavior than their constant flip angle counterparts and, being faster, should be less sensitive to patient motion, often seen in the longer TSE scans.

摘要

背景与目的

基于 TSE 的脊柱 T2 加权成像具有较长的扫描时间。本研究提出了一种使用可变重聚焦翻转角的快速成像方案,该方案针对模糊度和比吸收率进行了优化。

材料与方法

针对脊柱优化了可变重聚焦翻转角回波链,以改善点扩散函数并最小化比吸收率,与传统的恒定翻转角序列相比,生成的图像具有更高的空间分辨率和 SNR。使用常规 TSE 和所提出的序列(针对整个脊柱使用单次激发变体)从 51 名患者(35 名腰椎,16 名整个脊柱)中采集数据,进行非劣效性分析以评估所提出技术的效率。

结果

多shot 序列的扫描时间缩短了 2 倍,比吸收率降低了 >1.5 倍。与常规 TSE 相比,所提出的可变翻转角序列在所有图像质量和临床标准中均非劣效( <.025),除腰椎协议的信噪比外。然而,TSE-可变重聚焦翻转角的平均图像评分对于所有标准均≥4.3,一致性分析表明与 TSE 具有高度一致性(>90%),表明具有临床等效性。单次激发序列使整个脊柱的扫描时间缩短了 4 倍,并且所有标准的图像评分均≥4.4,证明了其临床实用性。

结论

我们提出了一种使用可变重聚焦翻转角的快速 T2 加权脊柱协议,包括单次激发变体。与恒定翻转角的序列相比,这些序列具有更好的点扩散函数行为,并且由于更快,应该对患者运动(在较长的 TSE 扫描中经常出现)的敏感性较低。

相似文献

本文引用的文献

4
Optimized three-dimensional fast-spin-echo MRI.优化的三维快速自旋回波磁共振成像
J Magn Reson Imaging. 2014 Apr;39(4):745-67. doi: 10.1002/jmri.24542. Epub 2014 Jan 8.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验