Pommé S, Paepen J, Van Ammel R
European Commission, Joint Research Centre (JRC), Directorate for Nuclear Safety and Security, Retieseweg 111, B-2440 Geel, Belgium.
European Commission, Joint Research Centre (JRC), Directorate for Nuclear Safety and Security, Retieseweg 111, B-2440 Geel, Belgium.
Appl Radiat Isot. 2018 Oct;140:171-178. doi: 10.1016/j.apradiso.2018.07.004. Epub 2018 Jul 6.
The half-life of Tc was measured at the JRC using the ionisation chamber 'IC1' (type Centronic IG12). The result, T(Tc) = 6.00660 (18) h, is in good agreement with literature data. One experiment was performed in IC1's default set-up with the ionisation current being integrated over an air capacitor and read out as a voltage increase over time. This ensured excellent linearity and precision throughout the dynamic range, but the maximum current was limited to 2 nA. In a second test, the current was directly read out with a Keithley 6517 A electrometer. Applying correction factors for the automatic range switching of the electrometer, an acceptable linearity was demonstrated over a range of 12 half-life periods starting at 20 nA. Range switching and autocorrelation of the current readout increase the systematic and random error propagation factors. Piecewise fitting of the decay curve over periods of 6 h yields the same Tc half-life value within 0.04% (0.0025 h) standard deviation over an activity range spanning at least 10 half-life periods (3 orders of magnitude).
在联合研究中心使用Centronic IG12型电离室“IC1”测量了锝的半衰期。结果为T(Tc) = 6.00660 (18) 小时,与文献数据吻合良好。在IC1的默认设置下进行了一项实验,电离电流在空气电容器上积分,并作为随时间增加的电压读出。这确保了在整个动态范围内具有出色的线性度和精度,但最大电流限制为2 nA。在第二项测试中,使用Keithley 6517 A静电计直接读出电流。应用静电计自动量程切换的校正因子,在从20 nA开始的12个半衰期范围内显示出可接受的线性度。电流读出的量程切换和自相关增加了系统误差和随机误差传播因子。在6小时的时间段内对衰变曲线进行分段拟合,在跨越至少10个半衰期(3个数量级)的活度范围内,得到的锝半衰期值在0.04%(0.0025小时)标准差内相同。