Chen Luguang, Liu Qi, Shi Zhang, Tian Xia, Peng Wenjia, Lu Jianping
Department of Radiology, Changhai Hospital of Shanghai, Second Military Medical University, Shanghai, China.
Diagn Interv Radiol. 2018 Jul;24(4):237-242. doi: 10.5152/dir.2018.17373.
We aimed to evaluate the interscan, intraobserver, and interobserver reproducibility of basilar atherosclerotic plaque employing dark blood high-resolution magnetic resonance imaging (HR-MRI) at 3 Tesla.
Sixteen patients (14 males and 2 females) with > 30% basilar stenosis as identified by conventional magnetic resonance angiography were prospectively recruited for scan and rescan examinations on a 3 Tesla MRI system using T2-weighted turbo spin-echo protocol. Two observers independently measured the areas of vessels and lumens. Wall area was derived by subtracting the lumen area from the vessel area. Areas of vessels, lumens and walls were compared for the evaluation of interscan variability of basilar plaque. To assess the intraobserver variability, one observer reevaluated all the images of the first scan after a 4-week interval.
Fourteen patients were included in the final analysis. No clinically significant difference was observed for interscan, intraobserver, and interobserver measurements. The intraclass correlations for vessel, lumen, and wall areas were excellent and ranged from 0.973 to 0.981 for the interscan measurements, 0.997 to 0.998 for the intraobserver measurements and 0.979 to 0.985 for the interobserver measurements. The coefficients of variation for quantitative basilar morphology measurements were 4.31%-10.35% for the interscan measurements, 1.41%-4.62% for the intraobserver measurements and 3.79%-8.46% for the interobserver measurements. Compared with the interscan and interobserver measurements, narrow intervals of the scatterplots were observed for the intraobserver measurements by Bland-Altman plots.
Basilar atherosclerotic plaque imaging demonstrates excellent reproducibility at 3 Tesla. The study proves that dark blood HR-MRI may serve as a reliable tool for clinical studies focused on the progression and treatment response of basilar atherosclerosis.
我们旨在评估使用3特斯拉场强的黑血高分辨率磁共振成像(HR-MRI)对基底动脉粥样硬化斑块进行扫描间、观察者内和观察者间的可重复性。
通过传统磁共振血管造影术确定有>30%基底动脉狭窄的16例患者(14例男性和2例女性)被前瞻性纳入,在3特斯拉MRI系统上使用T2加权快速自旋回波序列进行扫描和重新扫描检查。两名观察者独立测量血管和管腔的面积。壁面积通过从血管面积中减去管腔面积得出。比较血管、管腔和壁的面积,以评估基底动脉斑块的扫描间变异性。为评估观察者内变异性,一名观察者在间隔4周后重新评估第一次扫描的所有图像。
14例患者纳入最终分析。扫描间、观察者内和观察者间测量均未观察到临床显著差异。血管、管腔和壁面积的组内相关系数极佳,扫描间测量范围为0.973至0.981,观察者内测量为0.997至0.998,观察者间测量为0.979至0.985。基底动脉形态定量测量的变异系数扫描间测量为4.31%-10.35%,观察者内测量为1.41%-4.62%,观察者间测量为3.79%-8.46%。通过Bland-Altman图观察到,与扫描间和观察者间测量相比,观察者内测量的散点图区间更窄。
基底动脉粥样硬化斑块成像在3特斯拉场强下具有出色的可重复性。该研究证明黑血HR-MRI可作为专注于基底动脉粥样硬化进展和治疗反应的临床研究的可靠工具。