Suppr超能文献

用于采集氟化氢的硅胶吸附剂管的实验室评估

Laboratory evaluation of silica gel sorbent tubes for sampling hydrogen fluoride.

作者信息

Cassinelli M E

出版信息

Am Ind Hyg Assoc J. 1986 Apr;47(4):219-24. doi: 10.1080/15298668691389658.

Abstract

The NIOSH sampling and analytical method for inorganic acids employs a silica gel sorbent tube for the collection of five common inorganic acids with simultaneous determination of these acids in a single sample by ion chromatography. When the method was extended to the determination of hydrogen fluoride (HF) the sampled HF reacted with the silica gel and glass fiber of the sampler, but the reaction products remained trapped on the sorbent. Silica gel samplers were evaluated for the collection of HF from laboratory generated atmospheres. Factors tested included capacity, storage stability, humidity, accuracy and precision. Based on comparison with impinger collection, the mean recovery of HF from silica gel tube samples was 100.7% with a precision of 0.144.

摘要

美国国家职业安全与健康研究所(NIOSH)的无机酸采样与分析方法采用硅胶吸附管来采集五种常见无机酸,并通过离子色谱法在单个样品中同时测定这些酸。当该方法扩展到测定氟化氢(HF)时,采样的HF与采样器的硅胶和玻璃纤维发生反应,但反应产物仍被困在吸附剂上。对硅胶采样器在实验室产生的大气中采集HF的性能进行了评估。测试的因素包括容量、储存稳定性、湿度、准确性和精密度。与冲击式吸收管采集法相比,硅胶管样品中HF的平均回收率为100.7%,精密度为0.144。