Kalaiah Mohan Kumar, Lasrado Alina, Pinto Nikita, Shastri Usha
Department of Audiology and Speech Language Pathology, Kasturba Medical College, Mangalore, Manipal Academy of Higher Education, Manipal, Karnataka, India.
Department of Speech and Hearing Sciences, Arizona State University, Tempe, AZ, USA.
J Audiol Otol. 2018 Oct;22(4):189-196. doi: 10.7874/jao.2018.00038. Epub 2018 Aug 22.
Strength of medial olivocochlear reflex can be measured reliably using contralateral inhibition of distortion product otoacoustic emissions (DPOAEs) when its fine structure is considered. However, measurement of fine structure of DPOAE is difficult with clinical equipment. Thus, this study investigated the clinically relevant test-retest reliability of contralateral inhibition of DPOAEs. Subjects and.
Twenty-six young adults with normal hearing sensitivity participated. DPOAEs were recorded at 27 discrete f2 frequencies between 800 Hz and 8,000 Hz at frequency resolution of 8 points per octave with and without contralateral white noise presented at 50 dB SPL. To check for short term inter-session reliability, contralateral inhibition of DPOAEs were recorded in three sessions, two recording sessions on first day separated by 30 minutes and third time after one week of the first session. Within each session, DPOAEs were recorded twice in single probe-fit condition to test for intra-session reliability.
Cronbach's alpha was calculated having poor reliability (α≤ 0.7) of contralateral inhibition of DPOAEs in both intra-session and inter-session conditions for most of the tested frequencies. 95% confidence intervals of contralateral inhibition magnitude also showed large variability.
The current results showed that though DPOAE amplitudes were highly reliable across sessions, amount of inhibition of DPOAEs was not reliable when DPOAEs were measured at discrete frequencies. These findings are concurrent with the literature.
当考虑到内侧橄榄耳蜗反射的精细结构时,使用畸变产物耳声发射(DPOAE)的对侧抑制可以可靠地测量其强度。然而,使用临床设备测量DPOAE的精细结构很困难。因此,本研究调查了DPOAE对侧抑制在临床上相关的重测信度。对象与……
26名听力敏感度正常的年轻成年人参与研究。在800赫兹至8000赫兹之间的27个离散f2频率处记录DPOAE,频率分辨率为每倍频程8个点,分别在有和没有50分贝声压级的对侧白噪声的情况下进行记录。为了检查短期的不同时段信度,在三个时段记录DPOAE的对侧抑制,第一天的两个记录时段间隔30分钟,第三个时段在第一个时段一周后进行。在每个时段内,在单探头适配条件下记录DPOAE两次,以测试时段内信度。
对于大多数测试频率,在时段内和时段间条件下,计算得出的DPOAE对侧抑制的Cronbach's alpha信度较差(α≤0.7)。对侧抑制幅度的95%置信区间也显示出很大的变异性。
目前的结果表明,虽然DPOAE幅度在不同时段高度可靠,但在离散频率测量DPOAE时,其抑制量不可靠。这些发现与文献一致。