Effectiveness of 3-dimensional shoulder ultrasound in the diagnosis of rotator cuff tears: A meta-analysis.

作者信息

Teng Aiping, Liu Fanxiao, Zhou Dongsheng, He Tao, Chevalier Yan, Klar Roland M

机构信息

Department of Ultrasonography, Shandong Provincial Hospital Affiliated to Shandong University, Jinan, Shandong, China Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital of Munich (LMU), Munich, Germany Department of Orthopedics, Shandong Provincial Hospital Affiliated to Shandong University, Jinan, Shandong, China.

出版信息

Medicine (Baltimore). 2018 Sep;97(37):e12405. doi: 10.1097/MD.0000000000012405.

Abstract

BACKGROUND

Numerous quantitatively based studies measuring the accuracy of 3D shoulder ultrasound (US) for the diagnosis of rotator cuff tears remain inconclusive. In order to determine how effective 3D shoulder US is for detecting rotator cuff tears, a meta-analysis was performed systematically.

METHODS

Six electronic databases, PubMed/Medline, Embase, Cochrane Library, CNKI, VIP data, and Wanfang data, were utilized to retrieve articles praising the diagnostic value of 3D shoulder US for use in detecting rotator cuff tears. After screening and diluting out the articles that met inclusion criteria to be used for statistical analysis, the pooled evaluation indexes including sensitivity, specificity, and diagnostic odds ratio (DOR) as well as the summary receiver operating characteristic curve (SROC) were calculated utilizing Meta-Disc v.1.4.

RESULTS

Screening determined that out of 4220, 7 studies involving a total of 282 patients were deemed viable for inclusion in the meta-analysis. The results of the analysis showed that the sensitivity and specificity were at 94% and 83%, respectively, with a DOR of 60.06, Q* index of 0.9058 and the area under SROC of 0.9609. Additionally, a satisfactory accuracy of 3D shoulder US was observed in detecting full- and partial-thickness rotator cuff tears.

CONCLUSION

This meta-analysis suggests that 3D shoulder US is very effective and highly accurate to detect full-thickness rotator cuff tears, but may lack accuracy in the diagnosis of partial tears.

摘要

背景

众多基于定量分析的研究在测量三维肩部超声(US)诊断肩袖撕裂的准确性方面尚无定论。为了确定三维肩部超声在检测肩袖撕裂方面的有效性,我们系统地进行了一项荟萃分析。

方法

利用六个电子数据库,即PubMed/Medline、Embase、Cochrane图书馆、中国知网、维普数据和万方数据,检索赞扬三维肩部超声在检测肩袖撕裂中诊断价值的文章。在筛选并剔除符合纳入标准用于统计分析的文章后,使用Meta-Disc v.1.4计算合并评估指标,包括敏感性、特异性和诊断比值比(DOR)以及汇总接受者操作特征曲线(SROC)。

结果

筛选确定,在4220篇文章中,有7项研究共涉及282例患者被认为可纳入荟萃分析。分析结果显示,敏感性和特异性分别为94%和83%,DOR为60.06,Q*指数为0.9058,SROC下面积为0.9609。此外,在检测全层和部分厚度肩袖撕裂方面,三维肩部超声观察到了令人满意的准确性。

结论

这项荟萃分析表明,三维肩部超声在检测全层肩袖撕裂方面非常有效且高度准确,但在诊断部分撕裂方面可能缺乏准确性。