Although biotin interferences in TSH, FT3, FT4, and other biotinylated antibody-based assays manufactured by Roche Diagnostics have been well studied, there are relatively few reports on biotin interference in biotin-based assays manufactured by other companies. We investigated biotin interferences in TSH, FT4, and FT3 assays based on the LOCI (luminescent oxygen channeling assay) technology using the Dimension Vista 1500 analyzer (Siemens).
We prepared four serum pools using leftover specimens. Three serum pools were prepared initially for the original study but the 4 pool was prepared three months later. The aliquots of serum pool one and two were supplemented with various amounts of biotin (50 -1200 ng/mL) followed by determination of TSH, FT4, and FT3 concentrations. The aliquots of third pool were also supplemented with biotin to investigate whether 1:3 dilution could identify biotin interference. Aliquots of serum pool four were supplemented with biotin in order to study reproducibility of our original data.
We observed significantly elevated FT3 levels at biotin concentration of 100 ng/mL. In contrast, FT4 levels were falsely elevated but TSH levels were falsely decreased at a biotin level of 500 ng/mL. We also observed nonlinearity in dilution experiment.
We conclude that FT3 assay is most susceptible to biotin interference (threshold: 100 ng/mL) while the FT4 and TSH assays are less affected (threshold: 500 ng/mL). In addition, we also observed nonlinearity upon 1:3 dilution, which may indicate biotin interference (or interference from other compounds).
尽管罗氏诊断公司生产的促甲状腺激素(TSH)、游离三碘甲状腺原氨酸(FT3)、游离甲状腺素(FT4)及其他基于生物素化抗体的检测方法中的生物素干扰已得到充分研究,但关于其他公司生产的基于生物素的检测方法中的生物素干扰的报道相对较少。我们使用西门子公司的Dimension Vista 1500分析仪,对基于发光氧通道免疫分析(LOCI)技术的TSH、FT4和FT3检测中的生物素干扰进行了研究。
我们用剩余标本制备了四个血清混合样本。最初为原研究制备了三个血清混合样本,但第四个样本是在三个月后制备的。向血清混合样本一和二的等分试样中添加不同量的生物素(50 - 1200 ng/mL),随后测定TSH、FT4和FT3浓度。向第三个血清混合样本的等分试样中也添加生物素,以研究1:3稀释是否能识别生物素干扰。向血清混合样本四的等分试样中添加生物素,以研究我们原始数据的可重复性。
我们观察到,当生物素浓度为100 ng/mL时,FT3水平显著升高。相比之下,当生物素水平为500 ng/mL时,FT4水平出现假性升高,而TSH水平出现假性降低。我们在稀释实验中还观察到了非线性现象。
我们得出结论,FT3检测对生物素干扰最敏感(阈值:100 ng/mL),而FT4和TSH检测受影响较小(阈值:500 ng/mL)。此外,我们在1:3稀释时也观察到了非线性现象,这可能表明存在生物素干扰(或其他化合物的干扰)。